SU-E-J-174: ITV Variations as a Function of CT Geometry and Scan Time : A Simulation Study Using Patient Data
S St James1, P Mishra1, R Berbeco1
1Brigham and Women's Hospital, Boston, MA.
Increasing computed tomography (CT) scan duration can decrease internal target volume (ITV) uncertainties in lung cancer patients. Longer scan durations significantly impact ITV definition more than CT scanner geometry, suggesting modified CT protocols may improve treatment planning.
Area of Science:
- Medical Physics
- Radiotherapy
- Imaging Technology
Background:
- The internal target volume (ITV) is crucial for accounting for tumor motion and shape uncertainties in radiotherapy planning.
- Four-dimensional computed tomography (4DCT) is the standard imaging modality for ITV definition.
- Optimizing CT acquisition parameters is essential for accurate ITV delineation.
Purpose of the Study:
- To investigate the influence of CT scanner geometry and scan duration on internal target volume (ITV) definition.
- To utilize gold fiducial markers in lung tumors to represent tumor motion for simulation.
- To assess the impact of varying scan durations and CT geometries on ITV size.
Main Methods:
- Geometric simulations were performed in MATLAB, excluding image reconstruction variations.
- Simulated CT scanner geometries included a 4-slice scanner (10 mm axial field of view) and an 'ideal' scanner (100 mm axial field of view).
- Scan durations were varied from 6 to 30 seconds per bed position, with a 25 mm diameter spherical tumor model.
Main Results:
- Mean ITV consistently increased with increasing scan duration across all patients.
- The maximum observed increase in mean ITV was 25% with extended scan duration.
- Scan duration was identified as the primary factor influencing ITV size, surpassing scanner geometry effects.
Conclusions:
- Modified CT protocols, such as increasing scan duration, can potentially reduce ITV uncertainties.
- Decreasing scanner mA while increasing scan duration offers a strategy to manage radiation dose.
- Scan duration is a more significant determinant of ITV than CT scanner geometry, guiding protocol optimization.
More Related Videos
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies for Cardiovascular System V: CT
Imaging Studies III: Computed Tomography
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET