SU-E-I-29: Dose Rate and Image Quality Comparison Between Philips Veradius and GE OEC 9900 Elite Fluoroscopic C-Arms
A Springer1, E Solem1, R Kobistek1
1National Physics Consultants, Mentor, OH.
Purpose:
The radiation dose rate and image quality were compared on GE OEC 9900 Elite Vas 8 and Philips Veradius fluoroscopic C-arms.
Methods:
Plastic (PMMA) phantoms were used to simulate various thicknesses of soft tissues encountered during fluoroscopically-guided interventional procedures while aluminum and copper phantoms were used to simulate extremities of different sizes. An image quality test tool and an ionization chamber were included in the setup of each phantom. The phantoms were imaged in several fluoroscopy modes and at different dose levels. The entrance surface air KERMA rate (ESAKR) was measured for each phantom in each fluoroscopy mode. Simultaneously, the image quality was assessed by counting the number of low-contrast objects visible in the image quality test tool. For a given phantom type and thickness, the ESAKR was plotted versus the number of visible low-contrast objects to compare the dose rate required for a given level of image quality. Additionally, the mean ESAKR per number of objects was calculated and p-values were used to test for statistically significant differences between the C-arms.
Results:
For comparable image quality with a given phantom type and thickness the ESAKR with the OEC was typically much higher than with the Veradius. One-tailed, unpaired t-tests confirm that the mean ESAKR per number of visible low-contrast objects was lower for the Veradius than for the OEC with all thicknesses of PMMA (p < 0.05). The differences for the aluminum and copper phantoms were much smaller and were not statistically significant (p-values ranged from 0.2 to 0.47).
Conclusion:
In general, the Veradius C-arm provided better low-contrast image quality during fluoroscopy, with lower ESAKR than the OEC 9900 Elite Vas 8. The differences were statistically significant only with PMMA phantoms. Funding support provided by Philips Healthcare (Bothell, WA).
More Related Videos
06:59Improved Registration of 3D CT Angiography with X-ray Fluoroscopy for Image Fusion During Transcatheter Aortic Valve Implantation
Published on: June 3, 2018
06:50Troubleshooting FoCUS Image Acquisition: Patient Positioning, Transducer Manipulation, and Image Optimization
Published on: March 3, 2023
Related Concept Videos
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...
Imaging Studies VII: Vascular Imaging
Imaging Studies II: Ultrasonography