Test-Retest Reliability of the Dual-Microphone Voice Range Profile
Trine Printz1, Jesper Roed Sorensen1, Christian Godballe2
1Department of ORL Head & Neck Surgery, Odense University Hospital, Odense C, Denmark; Clinical Institute, University of Southern Denmark, Denmark.
Objectives:
The voice range profile (VRP) measures vocal intensity and fundamental frequency. Phonosurgical and logopedic treatment outcome studies using the VRP report voice improvements of 3-6 semitones (ST) in ST range and 4-7 decibels (dB) in sound pressure level range after treatment. These small improvements stress the importance of reliable measurements. The aim was to evaluate the test-retest reliability of the dual-microphone computerized VRP on participants with healthy voices.
Study Design:
This is a prospective test-retest reliability study.
Methods:
Dual-microphone VRPs were repeated twice on healthy participants (n = 37) with an interval of 6-37 days. Voice frequency and intensity (minimum, maximum, and ranges) were assessed in combination with the area of the VRP.
Results:
Correlations between VRP parameters were high (r > 0.60). However, in the retest, a statistically significant increase in voice frequency range (1.4 ST [95% confidence interval {CI}: 0.8-2.1 ST], P < 0.001), intensity ranges (2.2 dB [95% CI: 1.0-3.4 dB], P < 0.001), maximum frequency (1.0 ST [95% CI: 0.5-1.6 ST], P < 0.001), maximum intensity (1.4 dB [95% CI: 0.5-2.3 dB], P = 0.002), and area inside the VRP (148 cells [95% CI: 87-210 cells], P < 0.001) was observed.
Conclusion:
The intra-examiner variation of the dual-microphone VRP is well below the differences seen after surgical or logopedic intervention, even when measuring in non-sound-treated rooms. There is a need for studies regarding inter-examiner reliability with a longer interval between test and retest before the assessment is fully reliable for clinical application.
Related Concept Videos
Assessment of Ventilation II: Respiratory Depth and Rhythm
Respiratory depth measures the volume of air inhaled or exhaled during a breath. It can vary from shallow to deep and typically remains consistent when a person is at rest or asleep. Occasionally, individuals will automatically inhale deeply, known as sighing, which inflates the lungs with more air than normal breathing.
To assess respiratory depth, observe the degree of chest excursion or movement:
Instrument Calibration
Analytical Balance Calibration
An analytical balance measures mass and requires regular calibration to...
Reliability and Validity
Pulmonary Function Tests
Pulmonary Function Tests are crucial diagnostic tools for assessing respiratory function, particularly in patients with chronic respiratory disorders. They comprehensively evaluate lung volumes, ventilatory function, breathing mechanics, diffusion, and gas exchange. These tests help diagnose pulmonary diseases and play a significant role in monitoring disease progression, evaluating disability, and assessing response to therapy.
PFTs involve using a spirometer, a...
Root Mean Square
For example, consider the velocity of gas molecules in a container. The gas molecules are moving in different directions, which might impart positive and negative...