Related Experiment Video
Updated: Mar 2, 2026

Co-analysis of Brain Structure and Function using fMRI and Diffusion-weighted Imaging
Published on: November 8, 2012
EIT Imaging of admittivities with a D-bar method and spatial prior: experimental results for absolute and difference
1Department of Mathematics, Statistics, and Computer Science, Marquette University, Milwaukee, WI 53233, United States of America.
Abstract:
Electrical impedance tomography (EIT) is an emerging imaging modality that uses harmless electrical measurements taken on electrodes at a body's surface to recover information about the internal electrical conductivity and or permittivity. The image reconstruction task of EIT is a highly nonlinear inverse problem that is sensitive to noise and modeling errors making the image reconstruction task challenging. D-bar methods solve the nonlinear problem directly, bypassing the need for detailed and time-intensive forward models, to provide absolute (static) as well as time-difference EIT images. Coupling the D-bar methodology with the inclusion of high confidence a priori data results in a noise-robust regularized image reconstruction method. In this work, the a priori D-bar method for complex admittivities is demonstrated effective on experimental tank data for absolute imaging for the first time. Additionally, the method is adjusted for, and tested on, time-difference imaging scenarios. The ability of the method to be used for conductivity, permittivity, absolute as well as time-difference imaging provides the user with great flexibility without a high computational cost.