Extracranial internal carotid artery calcium volume measurement using computer tomography
Hediyeh Baradaran1, Chuen R Ng2, Ajay Gupta1
1Department of Radiology, Feil Family Brain and Mind Research Institute, Weill Cornell Medical College, New York, NY, USA.
Background:
The extent of calcium volume in the carotid arteries of contrast-based computer tomography (CT) is a valuable indicator of stroke risk. This study presents an automated, simple and fast calcium volume computation system. Since the high contrast agent can sometimes obscure the presence of calcium in the CT slices, it is therefore necessary to identify these slices before the corrected volume can be estimated.
Methods:
The system typically consists of segmenting the calcium region from the CT scan into slices based on Hounsfield Unit-based threshold, and subsequently computing the summation of the calcium areas in each slice. However, when the carotid volume has intermittently higher concentration of contrast agent, a dependable approach is adapted to correct the calcium region using the neighboring slices, thereby estimating the correct volume. Furthermore, mitigation is provided following the regulatory constraints by changing the system to semi-automated criteria as a fall back solution. We evaluate the automated and semi-automated techniques using completely manual calcium volumes computed based on the manual tracings by the neuroradiologist.
Results:
A total of 64 patients with calcified plaque in the internal carotid artery were analyzed. Using the above algorithm, our automated and semi-automated system yields correlation coefficients (CC) of 0.89 and 0.96 against first manual readings and 0.90 and 0.96 against second manual readings, respectively. Using the t-test, there was no significant difference between the automated and semi-automated methods against manual. The intra-observer reliability was excellent with CC 0.98.
Conclusions:
Compared to automated method, the semi-automated method for calcium volume is acceptable and closer to manual strategy for calcium volume. Further work evaluating and confirming the performance of our semi-automated protocol is now warranted.
More Related Videos
10:41Analysis of Cerebral Vasospasm in a Murine Model of Subarachnoid Hemorrhage with High Frequency Transcranial Duplex Ultrasound
Published on: June 3, 2021
08:12A Volumetric Method for Quantification of Cerebral Vasospasm in a Murine Model of Subarachnoid Hemorrhage
Published on: July 28, 2018
Related Concept Videos
Imaging Studies for Cardiovascular System VI: Calcium -Scoring CT
Imaging Studies for Cardiovascular System V: CT
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies III: Computed Tomography
Imaging Studies for Cardiovascular System IV: CMRI