Development and Testing of a Bubble Bi-Level Positive Airway Pressure System
Stephen C John1, Joseph D Barnett2, Nickolas D Habben2
1University of Michigan Medical School, Ann Arbor, Michigan. scjres@gmail.com.
Background:
Neonatal respiratory distress results in > 1 million annual deaths worldwide. Bubble CPAP is a simple, effective, and widely used therapy for infants in respiratory distress. In low-resource settings, more advanced respiratory support is limited by cost, technical expertise, and sporadic electricity. We sought to develop a safe, inexpensive, and simple solution to provide further respiratory support for these infants.
Methods:
A standard bubble CPAP system was modified to provide 2 levels of positive airway pressure (bi-level positive airway pressure) by attaching a novel device. To demonstrate reliability, the system was run with continuous pressure monitoring on full-term and preterm neonatal mannikins with pressure targets of 8/5 cm H2O and 15/5 cm H2O to simulate 2 different modes of noninvasive ventilation (NIV).
Results:
At a ventilation rate set between 30 and 45 cycles/min, by adjusting the leak rate of the device, the following mean pressures ± SD were demonstrated: term mannikin low-pressure NIV, 7.9 ± 0.2/5.3 ± 0.2 cm H2O; term mannikin high-pressure NIV, 15.1 ± 0.1/6.1 ± 0.1 cm H2O; preterm mannikin low-pressure NIV, 7.9 ± 0.2/5.3 ± 0.2 cm H2O; preterm mannikin high-pressure NIV, 16.5 ± 0.4/5.1 ± 0.1 cm H2O.
Conclusions:
The modified bubble CPAP system reliably provided alternating pressures similar to bi-level positive airway pressure modes of respiratory support in neonatal mannikins. The dual-pressure technology is a simple, single connection add-on that can readily be applied to existing bubble CPAP systems.
Related Concept Videos
Mechanical Ventilation III: Noninvasive Ventilation
Noninvasive Positive-Pressure Ventilation...
Cardiopulmonary Resuscitation II: ACLS Airway Management
Mechanical Ventilation I: Indication and Settings
Mechanical Ventilation II: Invasive Ventilation
Negative-Pressure Ventilators
Negative-pressure ventilators create a vacuum around the chest or body to draw air into the lungs, simulating breathing. This method does not require an...
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...
Ventilatory Modes
There are three ventilatory modes: full support, partial support, and spontaneous. These are described below.
Full Support Modes
Full support modes include controlled mechanical ventilation, continuous mandatory...