Performance Evaluation of Serum Free Light Chain Analysis: Nephelometry vs Turbidimetry, Monoclonal vs Polyclonal
Anne-Sophie Messiaen1, Maxime M W De Sloovere1, Paul-Emile Claus1
1From the Department of Laboratory Medicine, AZ Delta Roeselare-Menen, Roeselare, Belgium.
Objectives:
Free light chain (FLC) measurement gained a lot of interest for diagnostic workup of monoclonal gammopathy.
Methods:
We evaluated the performance of turbidimetric polyclonal Freelite (The Binding Site, Birmingham, UK) assays on Cobas 6000 (Roche Diagnostics, Rotkreuz, Switzerland) and nephelometric monoclonal N Latex (Siemens Healthcare Diagnostics, Marburg, Germany) assays on BN ProSpec (Dade Behring, Deerfield, IL) vs established nephelometric Freelite assays on BN ProSpec.
Results:
Analytical performance was acceptable. Method comparison (n = 118) showed significant proportional FLC differences for N Latex assays. However, good correlation and clinical concordance were shown. Recovery study in the low concentration range demonstrated consistent over- and underrecovery for Freelite reagents, hampering future research on prognostic value of suppressed noninvolved FLC. Antigen excess detection was successful for κ FLC in three-fourths of cases with Freelite reagents and in all cases with N Latex reagents. However, the latter resulted in underestimated κ FLC concentrations.
Conclusions:
FLC analysis requires continuous awareness of analytical limitations. Monitoring of disease response requires FLC analysis on the same platform using the same reagents.