Real-time electromagnetic tracking-based treatment platform for high-dose-rate prostate brachytherapy: Clinical
Luc Beaulieu1, Emmanuel Racine2, Dae Yup Han3
1Département de physique, de génie physique et d'optique et Centre de recherche sur le cancer de l'Université Laval, CHU de Québec, Québec, Québec, Canada; Département de radio-oncologie et Centre de recherche du CHU de Québec, CHU de Québec - Université Laval, Québec, Québec, Canada; Department of Radiation Oncology, University of California San Francisco, San Francisco.
Purpose:
New technologies were integrated into a novel treatment platform combining electromagnetically (EM) tracked catheters, a 3D ultrasound (3DUS) imaging device, and a new treatment planning system to provide a real-time prostate high-dose-rate (HDR) brachytherapy treatment system. This work defines workflows for offline CT and online 3DUS planning scenarios and preclinical end-to-end validation of the platform.
Methods And Materials:
The platform is composed of an EM-tracked stylet, a EM-tracked 3DUS probe, and an EM-tracked template guide, all used with the NDI Aurora field generator (NDI, Ontario, Canada). The treatment planning system performs continuous position and angular readings from all three EM sensors into a streamlined environment that allows for (1) contouring; (2) planning; (3) catheter insertion guidance and reconstruction; (4) QA of catheter path and tip position; and (5) exporting to an afterloader. Data were gathered on the times required for the various key steps of the 3DUS-based workflow.
Results:
The complete 3DUS-based workflow on 16-catheter implant phantoms took approximately 15 min. This time is expected to increase for actual patients. Plan generation is fast (7.6 ± 2.5s) and the initial catheter reconstruction with updated dose distribution is obtained at no (time) cost as part of the insertion process. Subsequent catheter reconstruction takes on average 10.5 ± 3.1s per catheter, representing less than 3 min for a 16-catheter implant.
Conclusions:
This preclinical study suggests that EM technology could help to significantly streamline real-time US-based high-dose-rate prostate brachytherapy.