Coronary Stent Failure: Fracture, Compression, Recoil, and Prolapse
Dominik M Wiktor1, Stephen W Waldo1, Ehrin J Armstrong1
1Division of Cardiology, VA Eastern Colorado Healthcare System, University of Colorado, Denver, CO, USA.
Abstract:
Current-generation coronary drug-eluting stents are associated with low rates of restenosis and target lesion revascularization. However, several mechanisms of stent failure remain clinically important. Stent fracture may occur in areas of excessive torsion or angulation. Longitudinal stent deformation is related to axial stent compression owing to extrinsic forces or secondary devices that disrupt stent architecture. Stent recoil occurs when a stent does not deploy at its optimal cross-sectional area. Tissue prolapse between stent struts may also predispose patients to adverse outcomes. Prevention, recognition, and treatment of these stent failures are necessary to optimize patient outcomes after percutaneous coronary interventions.