Precision and accuracy of consumer-grade motion tracking system for pedicle screw placement in pediatric spinal fusion surgery
Related Experiment Video
Updated: Feb 28, 2026

A Spine Robotic-Assisted Navigation System for Pedicle Screw Placement
Published on: May 11, 2020
Andrew Chan1, Janelle Aguillon2, Doug Hill3
1Department of Biomedical Engineering, University of Alberta, Edmonton, Alberta T6G 2V2, Canada.
Abstract:
Adolescent idiopathic scoliosis (AIS) is a 3-dimensional spinal deformity involving lateral curvature and axial rotation. Surgical intervention involves insertion of pedicle screws into the spine, requiring accuracies of 1mm and 5° in translation and rotation to prevent neural and vascular complications. While commercial CT-navigation is available, the significant cost, bulk and radiation dose hinders their use in AIS surgery. The objective of this study was to evaluate a commercial-grade Optitrack Prime 13W motion capture cameras to determine if they can achieve adequate accuracy for screw insertion guidance in AIS. Static precision, camera and tracked rigid body configurations, translational and rotational accuracy were investigated. A 1-h camera warm-up time was required to achieve precisions of 0.13mm and 0.10°. A three-camera system configuration with cameras at equal height but staggered depth achieved the best accuracy. A triangular rigid body with 7.9mm markers had superior accuracy. The translational accuracy for motions up to 150mm was 0.25mm while rotational accuracy was 4.9° for rotations in two directions from 0° to 70°. Required translational and rotational accuracies were achieved using this motion capture system as well as being comparable to surgical-grade navigators.