Intracorneal ring segment depth in keratoconus patients: a long-term follow-up study
Saied Shahhoseini1, Hassan Hashemi2, Soheila Asgari3
1Noor Ophthalmology Research Center, Noor Eye Hospital, Tehran, Iran.
International Ophthalmology
|June 14, 2017
Summary
Intracorneal ring segments (ICRS) implanted with femtosecond laser in keratoconus patients are placed shallower than intended. This finding suggests a need to recalibrate the femtosecond laser
Area of Science:
- Ophthalmology
- Corneal Surgery
- Laser Technology
Background:
- Keratoconus is a progressive corneal ectasia leading to irregular astigmatism.
- Intracorneal ring segments (ICRS) are used to regularize the cornea and improve visual acuity.
- Femtosecond laser technology offers precise tunnel creation for ICRS implantation.
Purpose of the Study:
- To compare the actual implantation depth of intracorneal ring segments (ICRS) created by femtosecond laser with the intended target depth.
- To evaluate the accuracy of femtosecond laser in achieving precise ICRS placement in keratoconus patients.
Main Methods:
- A retrospective chart review identified keratoconus patients who underwent femtosecond laser ICRS implantation.
- Anterior segment optical coherence tomography (AS-OCT) was used to measure the depth of the implanted ICRS.
- Depth was defined as the distance from the anterior corneal surface to the anterior ring segment.
Main Results:
- Thirty keratoconus patients with a mean follow-up of 25.8 months were analyzed.
- The mean actual ICRS depth was 59.9%, significantly shallower than the target depth of 85.0% (P < 0.001).
- Eighty-three percent (83.3%) of implanted rings were shallower than the intended 70% depth threshold.
Conclusions:
- Femtosecond laser-assisted ICRS implantation results in shallower placement than intended in keratoconus patients.
- Current femtosecond laser depth calculators may require revision to improve accuracy.
- Further investigation into optimizing femtosecond laser parameters for ICRS depth control is warranted.