A "quality-control-based correction method" for displayed dose indices on CT scanner consoles in patient dose surveys
Masoumeh Parsi1, Mehdi Sohrabi1, Fereidoun Mianji2
1Health Physics and Dosimetry Research Laboratory, Department of Energy Engineering and Physics, Amirkabir University of Technology, Tehran, Iran.
Purpose:
A new quality-control-based (QC-based) method is introduced to obtain correction factors to be applied to displayed patient dose indices (CTDIVol and DLP) on CT scanner consoles to verify improvement of dose surveys for diagnostic reference levels (DRLs) determination.
Method:
An available data-base of QC documents and reports of 57 CT scanners in Tehran, Iran was used to estimate CTDIVol, DLP and relevant correction factors for three CT examination types including head, chest and abdomen/pelvis. The correction factor is the ratio of QC-based estimated dose to displayed dose. A dose survey was performed by applying on-site "data collection method" and correction factors obtained in order to select CT scanners in three modes for determination of CT DRLs by inclusion of: (a) all CT scanners before displayed dose indices were corrected (57), (b) only CT scanners calibrated by QC experts (41) and (c) all CT scanners after displayed dose indices were corrected (57).
Results:
For the 41 CT scanners, correction factors of three examination types obtained in this study are within the acceptance tolerance of IAEA HHS-19. The correction factors range from 0.45 to 1.7 (average of 3 examinations) which is due to the change in the calibrated value of CTDIVol over extended time. The DRL differences in three modes are within ±1.0% for CTDIVol and ±12.4% for DLP.
Conclusions:
The "QC-based correction method" applied to mode (c) has improved the DRLs obtained by other two modes. This method is a strong alternative to "direct dose measurement" with simplicity and cost effectiveness.
More Related Videos
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Dose Response Curve: Conventional Versus Nonmonotonic
Imaging Studies III: Computed Tomography
Imaging Studies for Cardiovascular System VI: Calcium -Scoring CT
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET
Imaging Studies for Cardiovascular System V: CT