Retrospective Analysis of Continuous Glucose Monitoring Data With the Surveillance Error Grid Supports Nonadjunctive
John B Welsh1, Tomas Walker1, David Price1
11 Dexcom, Inc, San Diego, CA, USA.
Objective:
The objective was to assess clinical risks of inaccurate continuous glucose monitoring (CGM) system readings as estimated by the surveillance error grid (SEG).
Methods:
Values from Dexcom G4 Platinum system with an advanced algorithm (Software 505) were plotted on the SEG with temporally matched reference venous (YSI) values collected during clinic visits on days 1, 4, and 7 of sensor wear. Data from a pediatric study (N = 79, age [mean ± SD] 12.2 ± 4.6 years, all with type 1 diabetes) and an adult study (N = 51, age 46.7 ± 15.8 years, 44 with type 1 diabetes and 7 with type 2 diabetes) were used.
Results:
Pediatric data included 2262 paired points, of which 1990 (88.0%) were in the "no risk" zone. Adult data included 2263 paired points, of which 2056 (90.9%) were in the "no risk" zone. Performance was best on Day 4, when 92.7% and 93.3% of points from the pediatric and adult studies, respectively, were in the "no risk" zone. Nine of the 4525 points (<0.2%) from 5 different sensors were in zones representing moderate risk, and none were in zones representing great or extreme risk.
Conclusions:
SEG analysis suggests that in pediatric and adult subjects with diabetes, using CGM values for diabetes management poses minimal risk to the user. CGM users also benefit from glucose trends and alerts.
Related Concept Videos
Dosage Regimen Designs: Nomograms and Tabulations
Insulin: Dosing Regimen and Adverse Effects
The basal dose constitutes about 40%-50% of the total daily dose, with the rest as premeal insulin. The mealtime insulin dose should mirror...
Pharmacovigilance
This process, termed pharmacovigilance, aims to detect, evaluate, and minimize harmful effects related to medication use. The data collection for pharmacovigilance depends on spontaneous reporting systems, where healthcare professionals or patients voluntarily report suspected ADRs.
In some cases, there...
Dosage Regimen: Individualization
Dosage Regimens: Designs and Approaches
Hypoglycemia and Glucagon