Comparison between Bilistick System and transcutaneous bilirubin in assessing total bilirubin serum concentration in
C Greco1, I F Iskander2, D M Akmal2
1Bilimetrix s.r.l., AREA Science Park, Trieste, Italy.
Objective:
To compare the performance and accuracy of the JM-103 transcutaneous bilirubinometer and Bilistick System in measuring total serum bilirubin for the early identification of neonatal hyperbilirubinemia.
Study Design:
The study was performed on 126 consecutive term and near-term (⩾36 weeks' gestational age) jaundiced newborns in Cairo University Children Hospital NICU, Egypt. Total serum bilirubin was assayed concurrently by the clinical laboratory and Bilistick System and estimated using the JM-103 transcutaneous bilirubin instrument. Bland-Altman analysis was used to evaluate the agreement between determinations.
Result:
The limits of agreement of the Bilistick System (-5.8 to 3.3 mg dl-1) and JM-103 system (-5.4 to 6.0 mg dl-1) versus the clinical laboratory results were similar.
Conclusion:
The Bilistick System is an accurate alternative to transcutaneous (TcB) determination for early diagnosis and proper management of the neonatal jaundice.