Ultra-low radiation dose during electrophysiology procedures using optimized new generation fluoroscopy technology
James Crowhurst1,2, Haris Haqqani1,2, Daniel Wright1
1Cardiology Department, The Prince Charles Hospital, Chermside, Queensland, Australia.
Background:
Electrophysiology procedures require fluoroscopic guidance, with the associated potentially adverse effects of ionizing radiation. Newer fluoroscopy systems have more features that enable dose-reduction strategies. This study aimed to investigate any reduction in radiation dose between an older fluoroscopy system (Philips Integris H5000, Philips Healthcare, Einhoven, Netherlands) and one of the latest systems (Siemens Artis Q, Siemens Healthcare, Erlangen, Germany), optimized with dose-reduction strategies.
Methods:
Radiation dose measures were collected over a 2-year period in a single electrophysiology laboratory. Procedures were separated into seven groups: devices, biventricular devices, electrophysiology studies, standard radiofrequency ablation, complex atrial ablation, ablation for ventricular arrhythmias, and pulmonary vein isolation. In the first year, an older fluoroscopy system was used, and in the second year, a new system, with dose reduction strategies. Comparisons were also made to the literature with regard to radiation dose levels.
Results:
Patient characteristics, fluoroscopy times, number of digital acquisitions, procedural times, and procedural success were largely similar between the old and new system across procedure groups. Overall dose area product (DAP) was reduced by 91% (5.0 [2.0-17.0] to 0.45 [0.16-2.61] Gycm2 [P > 0.001]) with the new system and was lower across all groups. DAP readings with the new system are some of the lowest published in the literature in all groups.
Conclusion:
An optimized contemporary digital fluoroscopy system, with low radiation dose configuration and continued good procedural practice, can result in ultra-low radiation levels for all electrophysiology procedures, without compromising procedural time or procedural success.
More Related Videos
08:30X-ray Dose Reduction through Adaptive Exposure in Fluoroscopic Imaging
Published on: September 11, 2011
13:48Reduction of Radiation Exposure during Endovascular Treatment of Peripheral Arterial Disease Combining Fiber Optic RealShape Technology and Intravascular Ultrasound
Published on: April 21, 2023
Related Concept Videos
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...
Imaging Studies for Cardiovascular System V: CT