Is the Stella™ 5L system an effective cold sterilization technique for needle-based confocal miniprobes?
Cai-Xia Wang1, Yuan-Yuan Chen1, Feng Yang1
1Endoscopy Center, Shengjing Hospital of China Medical University, Shenyang 110004, Liaoning Province, China.
Background And Objective:
Needle-based confocal laser endomicroscopy (nCLE) under endoscopic ultrasound guidance is a newly developed imaging technique for pancreatic lesions and lymph nodes, enabling a subcellular level of resolution. The confocal miniprobe is an invasive instrument designed to be reused up to 10 times. Therefore, a method that ensures the complete elimination of microbial contaminants on the device is necessary. We studied the bactericidal efficacy of the Stella™ system, which purports to achieve this objective.
Materials And Methods:
The surfaces of three nCLE miniprobes were contaminated with suspensions of Bacillus atrophaeus (ATCC9372). One probe was randomly selected to count the bacterial load on the surface. The other two probes were manually cleaned and rinsed. One probe was randomly selected to count bacteria on the surface, and the other probe was sterilized using the Stella™ 5L endoscopic sterilization system before obtaining the bacterial count. The process was repeated for 20 cycles to evaluate the microbicidal efficacy of the Stella™ 5L endoscopic sterilization system. These miniprobes were immersed in the Stella Fuse disinfectant for 72 h. After the 72 h of immersion, the weight loss of probes was determined using a high precision electronic scale to examine corrosion following disinfection. The change in image quality was evaluated by an endoscopist.
Results:
From an initial contamination level of 4.48 × 106 ± 1.57 × 106 cfu/mL on the surface of the probes, the bacterial count was reduced to 4.25 × 102 ± 1.95 × 102 cfu/mL after manual cleaning (including enzyme washing), and no microorganisms were recovered after 20 cycles with the Stella™ 5L system. The probe weights before and after 72 h of immersion were 45.769 (45.768-45.771) g and 45.762 (45.752-45.768) g, respectively. No change in image quality was observed.
Conclusion:
This study shows that the Stella™ 5L system is capable of the complete elimination of microorganism contamination in a short period and avoids the toxicity of typical disinfectants. It is a safe, cheap, and efficient sterilization approach that provides a new option for nCLE miniprobe sterilization.
More Related Videos
Related Concept Videos
Methods of Sterilization I: Physical Methods
Steam sterilization uses non-toxic, low-cost moist heat in the form of saturated steam under pressure, which is fast, microbicidal, and sporicidal, and quickly warms and penetrates fabrics. Autoclaves, or steam sterilizers, expose each item to direct steam contact for a predetermined time at the necessary...