3D catheter reconstruction in HDR prostate brachytherapy for pre-treatment verification using a flat panel detector
Ryan L Smith1, Annette Haworth2, Vanessa Panettieri3
1Alfred Health Radiation Oncology, The Alfred Hospital, Melbourne 3004, VIC, Australia; School of Science, RMIT University, Melbourne 3000, VIC, Australia.
Purpose:
High dose rate prostate brachytherapy is a widely-practiced treatment, delivering large conformal doses in relatively few treatment fractions. Inter- and intra-fraction catheter displacements have been reported. Unrecognized displacement can have a significant impact on dosimetry. Knowledge of the implant geometry at the time of treatment is important for ensuring safe and effective treatment. In this work we demonstrate a method to reconstruct the catheter positions pre-treatment, using a 'shift' imaging technique, and perform registration with the treatment plan for verification relative to the prostate.
Methods:
Two oblique 'shift' images were acquired of a phantom containing brachytherapy catheters, representing the patient immediately pre-treatment. Using a back projection approach, the catheter paths were reconstructed in 3D and registered with the planned catheter paths. The robustness of the reconstruction and registration process was investigated as a function of phantom rotation. Catheter displacement detection was performed and compared to known applied displacements.
Results:
Reconstruction of the implant geometry in 3D immediately prior to treatment was achieved. A mean reconstruction uncertainty of 0.8mm was determined for all catheters with a mean registration uncertainty of 0.5mm. A catheter displacement detection threshold of 2.2mm was demonstrated. Catheter displacements were all detected to within 0.5mm of the applied displacements.
Conclusion:
This technique is robust and sensitive to assess catheter displacements throughout the implant volume. This approach provides a method to detect, in 3D, changes in catheter positions relative to the prostate. The method has sufficient sensitivity to enable clinically significant decisions immediately prior to treatment delivery.