The effect of concomitant fields in fast spin echo acquisition on asymmetric MRI gradient systems
Shengzhen Tao1,2, Paul T Weavers1, Joshua D Trzasko1
1Department of Radiology, Mayo Clinic, Rochester, Minnesota, USA.
Purpose:
To investigate the effect of the asymmetric gradient concomitant fields (CF) with zeroth and first-order spatial dependence on fast/turbo spin-echo acquisitions, and to demonstrate the effectiveness of their real-time compensation.
Methods:
After briefly reviewing the CF produced by asymmetric gradients, the effects of the additional zeroth and first-order CFs on these systems are investigated using extended-phase graph simulations. Phantom and in vivo experiments are performed to corroborate the simulation. Experiments are performed before and after the real-time compensations using frequency tracking and gradient pre-emphasis to demonstrate their effectiveness in correcting the additional CFs. The interaction between the CFs and prescan-based correction to compensate for eddy currents is also investigated.
Results:
It is demonstrated that, unlike the second-order CFs on conventional gradients, the additional zeroth/first-order CFs on asymmetric gradients cause substantial signal loss and dark banding in fast spin-echo acquisitions within a typical brain-scan field of view. They can confound the prescan correction for eddy currents and degrade image quality. Performing real-time compensation successfully eliminates the artifacts.
Conclusions:
We demonstrate that the zeroth/first-order CFs specific to asymmetric gradients can cause substantial artifacts, including signal loss and dark bands for brain imaging. These effects can be corrected using real-time compensation. Magn Reson Med 79:1354-1364, 2018. © 2017 International Society for Magnetic Resonance in Medicine.
Related Concept Videos
Magnetic Resonance Imaging
Atomic Nuclei: Magnetic Resonance
NMR Spectrometers: Radiofrequency Pulses and Pulse Sequences
Atomic Nuclei: Nuclear Relaxation Processes
NMR Spectroscopy: Spin–Spin Coupling
¹H NMR: Interpreting Distorted and Overlapping Signals
As Δν decreases and the signals move closer, the doublets appear increasingly distorted. The intensities of the inner lines increase at the cost of those of the outer lines as the signals are...