Wavelength-specific lighted suction instrument for 5-aminolevulinic acid fluorescence-guided resection of deep-seated malignant glioma: technical note
Related Experiment Video
Updated: Feb 27, 2026

Image-Guided Resection of Glioblastoma and Intracranial Implantation of Therapeutic Stem Cell-seeded Scaffolds
Published on: July 16, 2018
Abstract:
Surgery guided by 5-aminolevulinic acid (ALA) fluorescence has become a valuable adjunct in the resection of malignant intracranial gliomas. Furthermore, the fluorescence intensity of biopsied areas of a resection cavity correlates with histological identification of tumor cells. However, in the case of lesions deep within a resection cavity, light penetration may be suboptimal, resulting in less excitation of 5-ALA metabolites, leading to decreased fluorescence emission. To address this obstacle, the authors report on the use of a 400-nm wavelength fiber-optic lighted suction instrument that can be used both during resection of a tumor and to provide direct light to deeper areas of a resection cavity. In the presented case, this wavelength-specific lighted suction instrument improved the fluorescence intensity of patches of malignant tissue within the resection cavity. This technique may further improve the utility of 5-ALA in identifying tumor-infiltrated tissue for deep-seated lesions. Additionally, this tool may have implications for scoring systems that correlate 5-ALA fluorescence intensity with histological identification of malignant cells.