Related Experiment Video
Updated: Feb 27, 2026

Highly Sensitive and Quantitative Detection of Proteins and Their Isoforms by Capillary Isoelectric Focusing Method
Published on: September 19, 2018
Quantitation of CRM197 using imaged capillary isoelectric focusing with fluorescence detection and capillary Western
John W Loughney1, Sha Ha1, Richard R Rustandi1
1Biologics and Vaccines, Merck Research Laboratories, Merck and Co., Inc., Kenilworth, NJ, USA.
Abstract:
Maurice is a new instrument that can perform imaged capillary isoelectric focusing (icIEF). The standard detection for icIEF is UV absorbance at 280 nm, which limits its application to high protein concentration samples and non-complex samples. Here we describe an icIEF instrument with fluorescence detection. We demonstrate the advantage of using either icIEF with fluorescence detection or quantitative Western Blot to measure diphtheria toxin mutant CRM197 protein titer in crude cell lysates and purified samples. These two techniques have great potentials to become standard methods to analyze protein titers in crude cell lysate or other complex samples types.
Related Concept Videos
Capillary Electrophoresis: Applications
Capillary zone electrophoresis (CZE) separates ionic components based on their electrophoretic mobility. It has been used to separate proteins, amino acids,...
Electrophoresis: Overview
There...