Measuring the Carotid to Femoral Pulse Wave Velocity Cf-PWV to Evaluate Arterial Stiffness
Published on: May 3, 2018
Camera-based assessment of arterial stiffness and wave reflection parameters from neck micro-motion
A V Moço1, L Z Mondragon1, W Wang1
1Department of Electrical Engineering, Eindhoven University of Technology, 5612AZ Eindhoven, Netherlands.
Abstract:
The feasibility of a camera-based extraction of carotid distension waveforms offers the prospect of a user-friendly alternative to laser Doppler velocimetry (LDV) or accelerometry-based systems. Upon supplementary calibration of vessel wall displacement to arterial pressure, our system may also be an appealing alternative to applanation tonometry for extracting cardiac-related features from the central pulse pressure waveform.
Objective:
This paper describes the application of camera-based micro-motion imaging to extract health-related features from the contour of the carotid displacement waveform.
Approach:
We build on the assumption that the cardiac-related frequency components of the skin motion (sMOT) waveform, as acquired at the vicinity of the carotid artery under uneven illumination, receive a dominant contribution from the carotid wall displacement. We propose a two-step approach at which sMOT signals are queried based on the local amplitude of remote-photoplethysmography sensors spanning the neck's skin and then ensemble-averaged for cardiovascular health assessment.
Main Results:
The feasibility of the system is demonstrated for assessing stiffness index, augmentation pressure, augmentation index and reflection magnitude on a dataset comprising 28 participants (ages 23 to 62 yrs; 22 males).
Significance:
Although presented here as a standalone system, micro-motion imaging can be an auxiliary technique for improving sensor placement and signal quality of tonometric or LDV technologies.
Related Concept Videos
Assessment of blood pressure in brachial artery(one-step method)
Prepare for the Procedure:
Assessment of blood pressure in brachial artery(two-step method)
Assessing Blood pressure using a doppler ultrasound
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment:
Assessment of apical radial pulse
The A-R pulse assessment involves simultaneous evaluation of the apical and radial pulses. When the apical and radial pulse rates vary, this assessment helps identify a pulse deficit.
Pre-Procedural Preparation
Assessment of radial pulse
The radial pulse, located at the wrist, is often the preferred site for assessing peripheral pulse because of its accessibility and dependability. The process of determining the radial pulse involves several steps:
Assessing Blood pressure in the Leg
Preparation: