Performance of automated digital cell imaging analyzer Sysmex DI-60
Hyeong Nyeon Kim1, Mina Hur1, Hanah Kim1
1.
The Sysmex DI-60 automated analyzer shows acceptable performance for white blood cell differentials and platelet estimation, comparable to manual counts. Verification improves red blood cell classification accuracy, optimizing hematology lab workflows.
Area of Science:
- Hematology
- Clinical Pathology
- Medical Laboratory Technology
Background:
- The Sysmex DI-60 is a new automated digital cell imaging analyzer.
- Evaluating its performance against established methods is crucial for clinical adoption.
Purpose of the Study:
- To assess the diagnostic performance of the Sysmex DI-60 analyzer.
- To compare its accuracy in white blood cell (WBC) differentials, red blood cell (RBC) classification, and platelet (PLT) estimation against the Sysmex XN analyzer and manual counts.
Main Methods:
- A total of 276 samples (176 abnormal, 100 normal) were analyzed.
- WBC differentials, RBC classification, and PLT estimation were performed using DI-60 and compared with Sysmex XN and/or manual counts.
- RBC morphology and manual counts followed ICSH and CLSI guidelines, respectively.
Main Results:
- Overall concordance for WBCs between DI-60 and manual count was 86.0%.
- Excellent agreement (weighted κ=0.963) was observed for WBC five-part differentials after verification.
- Strong correlations were found for neutrophils (r=0.955), lymphocytes (r=0.871), immature granulocytes (r=0.820), and blasts (r=0.879).
- Platelet counts by DI-60 highly correlated with XN (r=0.945), though it underestimated counts in marked thrombocytosis.
- Notable differences were observed in RBC grading between DI-60 and manual counting based on ICSH criteria.
Conclusions:
- The Sysmex DI-60 demonstrates acceptable performance for WBC differential, RBC classification, and platelet estimation, particularly after verification.
- The system has the potential to optimize hematology laboratory workflows by reducing manual workload.