Shear-Dependent Agreement and Clinical Reclassification of Whole-Blood Viscosity Measurements: A Paired Comparison of
Jongho Yi1, Hong-Geun Jung2, Seoung Joon Lee2
1Department of Laboratory Medicine, Konkuk University School of Medicine, Seoul 05030, Republic of Korea.
Background/Objectives: Whole-blood viscosity (WBV) is increasingly used in cardiovascular risk assessment; however, inter-device comparability may depend on shear-rate definition. We performed a paired comparison of two scanning capillary viscometers to evaluate shear-dependent analytical agreement and its impact on clinical classification. Methods: In 300 identical blood samples, WBV was measured using Rheovis 2000A and Hemovister. Systolic WBV was defined at 300 s-1 for both devices (shear-matched), whereas clinically defined diastolic WBV corresponded to 1 s-1 for Rheovis 2000A and 5 s-1 for Hemovister. Agreement was assessed using linear regression and Bland-Altman analysis. Hematocrit tertiles were examined as effect modifiers. Clinical agreement was evaluated using quadratic weighted Cohen's κ. Results: Across matched shear rates (1000 to 1 s-1), Hemovister yielded consistently higher WBV values than Rheovis 2000A, with statistically significant inter-device differences at all shear levels except 1000 s-1. The magnitude of bias increased progressively as shear rate decreased, reaching -8.34 mPa·s at 1 s-1. Under shear-matched systolic conditions (300 s-1), the mean difference was -0.25 mPa·s (limits of agreement -1.72 to 1.22). In contrast, under clinically defined diastolic conditions (1 vs. 5 s-1), the mean difference was 14.54 mPa·s (3.93 to 25.15), increasing across hematocrit tertiles. Clinical agreement was fair for systolic (κ = 0.31; 95% CI 0.24 to 0.39) and moderate for diastolic WBV (κ = 0.44; 95% CI 0.37 to 0.51). Notably, among samples classified as high by Hemovister, 72.8% (systolic) and 54.0% (diastolic) were reclassified as normal by Rheovis 2000A. Conclusions: Inter-device agreement in WBV measurement is strongly shear-dependent. Although numerical divergence increases at low shear, categorical concordance may remain moderate when device-specific reference thresholds are applied. Harmonization of shear definitions and reference frameworks may therefore be essential for consistent cross-platform interpretation.
Background/Objectives: Whole-blood viscosity (WBV) is increasingly used in cardiovascular risk assessment; however, inter-device comparability may depend on shear-rate definition. We performed a paired comparison of two scanning capillary viscometers to evaluate shear-dependent analytical agreement and its impact on clinical classification. Methods: In 300 identical blood samples, WBV was measured using Rheovis 2000A and Hemovister. Systolic WBV was defined at 300 s-1 for both devices (shear-matched), whereas clinically defined diastolic WBV corresponded to 1 s-1 for Rheovis 2000A and 5 s-1 for Hemovister. Agreement was assessed using linear regression and Bland-Altman analysis. Hematocrit tertiles were examined as effect modifiers. Clinical agreement was evaluated using quadratic weighted Cohen's κ. Results: Across matched shear rates (1000 to 1 s-1), Hemovister yielded consistently higher WBV values than Rheovis 2000A, with statistically significant inter-device differences at all shear levels except 1000 s-1. The magnitude of bias increased progressively as shear rate decreased, reaching -8.34 mPa·s at 1 s-1. Under shear-matched systolic conditions (300 s-1), the mean difference was -0.25 mPa·s (limits of agreement -1.72 to 1.22). In contrast, under clinically defined diastolic conditions (1 vs. 5 s-1), the mean difference was 14.54 mPa·s (3.93 to 25.15), increasing across hematocrit tertiles. Clinical agreement was fair for systolic (κ = 0.31; 95% CI 0.24 to 0.39) and moderate for diastolic WBV (κ = 0.44; 95% CI 0.37 to 0.51). Notably, among samples classified as high by Hemovister, 72.8% (systolic) and 54.0% (diastolic) were reclassified as normal by Rheovis 2000A. Conclusions: Inter-device agreement in WBV measurement is strongly shear-dependent. Although numerical divergence increases at low shear, categorical concordance may remain moderate when device-specific reference thresholds are applied. Harmonization of shear definitions and reference frameworks may therefore be essential for consistent cross-platform interpretation.
Related Concept Videos
Blood Transfusion and Agglutination
History
The history of blood transfusion dates back to the 17th century, when early attempts were made in animals. In 1818 James Blundell, a British doctor, performed the first successful human blood transfusion. Later in 1900, Karl...
Bioequivalence studies: Biowaivers