Multimodality Diagnosis of Mesenteric Ischemia
Published on: July 21, 2023
Gastrointestinal Impedance Spectroscopy to Detect Hypoperfusion During Hemorrhage
Andreas Bloch1, Andreas Kohler, Horst Posthaus
1*Department of Intensive Care Medicine, Inselspital, Bern University Hospital, University of Bern, Bern, Switzerland†Department for Visceral Surgery and Medicine, Inselspital, Bern University Hospital, University of Bern, Bern, Switzerland‡Institute for Animal Pathology, Vetsuisse Faculty, University of Bern, Bern, Switzerland§Helbling Technik Bern AG, Liebefeld-Bern, Bern, Switzerland.
Gastrointestinal impedance spectroscopy did not detect early changes in tissue perfusion during hemorrhage or retransfusion in pigs. This technique is unlikely to provide additional information on hypovolemia-induced gastrointestinal perfusion changes.
Area of Science:
- Physiology
- Medical Devices
- Surgical Research
Background:
- Tissue impedance changes are potential early indicators of impaired tissue perfusion.
- Hemorrhage can lead to compromised blood flow in the gastrointestinal tract.
- Impedance spectroscopy is a technique that measures electrical properties of tissues.
Purpose of the Study:
- To evaluate the utility of impedance spectroscopy in detecting early changes in alimentary tract tissue impedance during acute hemorrhage.
- To determine if impedance spectroscopy can identify alterations in gastrointestinal perfusion caused by blood loss and retransfusion.
Main Methods:
- Acute hemorrhage was induced in anesthetized pigs, with stepwise reductions in mean arterial blood pressure (MAP).
- Tissue impedance was measured at multiple gastrointestinal sites (sublingual, esophagus, stomach, proximal jejunum).
- Simultaneously, cardiac output, superior mesenteric artery blood flow, and jejunal mucosal blood flow were monitored.
Main Results:
- Despite significant hypotension and reduced blood flow, impedance spectroscopy showed no changes in tissue impedance at any monitored site during hemorrhage or retransfusion.
- Enteral nutrition administration did not affect tissue impedance or blood flow measurements.
- Histological examination revealed focal gastric necrosis in some animals at the impedance transducer site.
Conclusions:
- Gastrointestinal impedance spectroscopy is not effective in detecting early signs of impaired tissue perfusion during acute hemorrhage and retransfusion.
- The technique offers limited value for assessing hypovolemia-induced changes in gastrointestinal blood flow.