Related Experiment Video
Updated: Feb 26, 2026

Targeting Neuronal Fiber Tracts for Deep Brain Stimulation Therapy Using Interactive, Patient-Specific Models
Published on: August 12, 2018
Systematic review of hardware-related complications of Deep Brain Stimulation: Do new indications pose an increased
Onanong Jitkritsadakul1, Roongroj Bhidayasiri2, Suneil K Kalia3
1Morton and Gloria Shulman Movement Disorders Centre and the Edmond J. Safra Program in Parkinson's Disease, Toronto Western Hospital, UHN, Toronto, Ontario, Canada; Chulalongkorn Center of Excellence for Parkinson's Disease & Related Disorders, Department of Medicine, Faculty of Medicine, Chulalongkorn University and King Chulalongkorn Memorial Hospital, Thai Red Cross Society, Bangkok, Thailand.
Introduction:
Deep Brain Stimulation (DBS) is an effective treatment extended broadly to many neurological and psychiatric disorders. Nevertheless, complications may arise during DBS procedures or following implantation due to implanted hardware. This may result in both minor and major adverse events that may necessitate hardware removal and/or compromise maximal therapeutic benefit for the patient.
Objectives And Methods:
To identify relevant literature on hardware-related complications from DBS procedures by performing a systematic review, and propose how to identify at-risk group and possible preventive approaches.
Results:
Of 4592 abstract screened, 96 articles fulfilled the selection criteria and were reviewed. Overall, the most common hardware-related complications were infections (5.12% of patients), followed by lead migration (1.60%), fracture or failure of the lead or other parts of the implant (1.46% and 0.73%, respectively), IPG malfunctions (1.06% of patients), and skin erosions without infections (0.48% of patients). New indications for DBS, including Tourette's syndrome, cluster headache, and refractory partial epilepsy, were found to bear a higher incidence of hardware-related infections than established indications such as Parkinson's disease. The highest rate of lead fracture or failure was found in dystonia patients (4.22%). Ultimately, the highest rate of pain at the implantation sites was found in refractory partial epilepsy patients (16.55%).
Conclusion:
Our analysis identified a variety of potential hardware-related complications among patients who underwent DBS procedures. Patients who were at risk of complications, such as patients with dystonia and off-label indications (e.g. Tourette's syndrome) should be informed prior to surgery and closely followed thereafter.