Validation of a sheet-shaped body vibrometer for screening of obstructive sleep apnea
Takamasa Kogure1,2, Mina Kobayashi3,4, Takashi Okawa5
1Department of Somnology, Tokyo Medical University.
Abstract:
We assessed the validity of using a sheet-shaped body vibrometer (SBV) as a portable monitoring device for obstructive sleep apnea (OSA) screening. Seventy consecutive patients with suspected OSA underwent simultaneous in-laboratory polysomnography (PSG) and SBV. We evaluated the screening accuracy of the respiratory event index (REI) obtained with the SBV, using the REI based on either the estimated total sleep time (REI_eTST) or time in bed (REI_TIB); these were compared to the apnea-hypopnea index (AHI) obtained via PSG. Bland-Altman plots indicated that the mean difference between REI_eTST and AHI was lower than that between REI_TIB and AHI (1.2 ± 19.8 vs. 6.5 ± 16.8). For AHI ≥ 15, the sensitivity and specificity at an optimal REI_eTST of 17.0 were 90.9% and 76.9%, whereas those at an optimal REI_TIB of 15.9 were 86.4% and 80.8%, respectively; moreover, for AHI ≥ 30, these values at an optimal REI_eTST of 26.0 were 89.5% and 88.2%, whereas those at an optimal REI_TIB of 23.8 were 84.2% and 92.2%, respectively. The optimal cutoff values of REIs for AHI of ≥ 5 were markedly different from those for AHI obtained via PSG (REI_eTST, 14.9; REI_TIB, 15.0), but close to those for AHI of ≥ 15; both had good sensitivities and specificities. REIs obtained via SBV performed well in moderate-to-severe, but not mild, OSA screening; REI_eTST showed a slightly higher sensitivity and a relatively closer value to the AHI obtained via PSG when compared to REI_TIB. We consider the SBV less acceptable for screening mild cases than more severe cases.
Related Concept Videos
Sleep Apnea
The condition is more prevalent among...
Physical Assessment of the Respiratory Tract III: Percussion
Percussion in Respiratory Assessment
Percussion evaluates underlying tissue composition with audible and tactile vibrations,...
Assessment of Ventilation II: Respiratory Depth and Rhythm
Respiratory depth measures the volume of air inhaled or exhaled during a breath. It can vary from shallow to deep and typically remains consistent when a person is at rest or asleep. Occasionally, individuals will automatically inhale deeply, known as sighing, which inflates the lungs with more air than normal breathing.
To assess respiratory depth, observe the degree of chest excursion or movement:
Respiratory System Abnormal Finding I: Inspection and Percussion
Inspection Findings
During an inspection, several findings may suggest the presence of respiratory distress or disease. Pursed-lip breathing, where exhalation is slowed by...
Physical Assessment of the Respiratory Tract II: Palpation
Thoracic Palpation
Thoracic palpation detects tenderness, masses, lesions, respiratory excursions, and vocal fremitus. The nurse assesses...
Respiratory System Abnormal Finding II: Palpation and Auscultation
Palpation Findings
During a respiratory assessment, palpation can reveal several vital abnormalities: