Related Experiment Video
Updated: Feb 25, 2026

Gradient Echo Quantum Memory in Warm Atomic Vapor
Published on: November 11, 2013
Pulsed high magnetic field measurement with a rubidium vapor sensor
S George1, N Bruyant1, J Béard1
1Laboratoire National des Champs Magnétiques Intenses (UPR 3228, CNRS-UPS-UGA-INSA), F-31400 Toulouse Cedex, France.
Abstract:
We present a new technique to measure pulsed magnetic fields based on the use of rubidium in gas phase as a metrological standard. We have therefore developed an instrument based on laser inducing transitions at about 780 nm (D2 line) in rubidium gas contained in a mini-cell of 3 mm × 3 mm cross section. To be able to insert such a cell in a standard high-field pulsed magnet, we have developed a fibred probe kept at a fixed temperature. Transition frequencies for both the π (light polarization parallel to the magnetic field) and σ (light polarization perpendicular to the magnetic field) configurations are measured by a commercial wavemeter. One innovation of our sensor is that in addition to the usual monitoring of the light transmitted by the Rb cell, we also monitor the fluorescence emission of the gas sample from a volume of 0.13 mm3. Our sensor has been tested up to about 58 T.
Related Concept Videos
NMR Spectrometers: Radiofrequency Pulses and Pulse Sequences
Atomic Nuclei: Magnetic Resonance
Atomic Nuclei: Nuclear Relaxation Processes
NMR Spectrometers: Overview