Technical note: TROG 15.01 SPARK trial multi-institutional imaging dose measurement
Kimberley Legge1, Peter B Greer1,2, Paul J Keall3
1School of Mathematical and Physical Sciences, University of Newcastle, Callaghan, New South Wales, Australia.
Purpose:
The Trans-Tasman Radiation Oncology Group (TROG) 15.01 Stereotactic Prostate Adaptive Radiotherapy utilizing Kilovoltage intrafraction monitoring (SPARK) trial is a multicenter trial using Kilovoltage Intrafraction Monitoring (KIM) to monitor prostate position during the delivery of prostate radiation therapy. KIM increases the accuracy of prostate radiation therapy treatments and allows for hypofractionation. However, an additional imaging dose is delivered to the patient. A standardized procedure to determine the imaging dose per frame delivered using KIM was developed and applied at four radiation therapy centers on three different types of linear accelerator.
Methods:
Dose per frame for kilovoltage imaging in fluoroscopy mode was measured in air at isocenter using an ion chamber. Beam quality and dose were determined for a Varian Clinac iX linear accelerator, a Varian Trilogy, four Varian Truebeams and one Elekta Synergy at four different radiation therapy centers. The imaging parameters used on the Varian machines were 125 kV, 80 mA, and 13 ms. The Elekta machine was measured at 120 kV, 80 mA, and 12 ms. Absorbed doses to the skin and the prostate for a typical SBRT prostate treatment length were estimated according to the IPEMB protocol.
Results:
The average dose per kV frame to the skin was 0.24 ± 0.03 mGy. The average estimated absorbed dose to the prostate for all five treatment fractions across all machines measured was 39.9 ± 2.6 mGy for 1 Hz imaging, 199.7 ± 13.2 mGy for 5 Hz imaging and 439.3 ± 29.0 mGy for 11 Hz imaging.
Conclusions:
All machines measured agreed to within 20%. Additional dose to the prostate from using KIM is at most 1.3% of the prescribed dose of 36.25 Gy in five fractions delivered during the trial.
More Related Videos
10:33Expedited Radiation Biodosimetry by Automated Dicentric Chromosome Identification ADCI and Dose Estimation
Published on: September 4, 2017
06:20Irradiator Commissioning and Dosimetry for Assessment of LQ α and β Parameters, Radiation Dosing Schema, and in vivo Dose Deposition
Published on: March 11, 2021
Related Concept Videos
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET
Dose Size and Dosing Frequency: Determination Methods
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...
Imaging Studies III: Computed Tomography
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...