Silent navigator-triggered silent MRI of the abdomen
Yuji Iwadate1, Atsushi Nozaki1, Yoshinobu Nunokawa2
1Global MR Applications and Workflow, GE Healthcare Japan, Hino, Tokyo, Japan.
Purpose:
To develop and demonstrate the feasibility of a silent respiratory navigator technique for prospective triggering, which was incorporated into a three-dimensional radial zero-echo-time sequence for respiratory navigated silent abdominal imaging.
Methods:
A nonselective hard excitation radiofrequency pulse was used for the navigator sequence with a derated readout gradient, to avoid generation of high levels of acoustic noise. The acquired navigator signals were processed in real time and used for prospective triggering of the zero-echo-time sequence. Ten healthy volunteers were scanned using the proposed and conventional techniques at 1.5 T. An acoustic noise measurement with A-weighted continuous equivalent sound pressure level was also performed.
Results:
The sound pressure-level values of the background noise, zero-echo-time imaging, conventional, and silent navigators were 68.3, 68.4, 102.5, and 69.4 dBA, respectively. Excellent correlation with correlation coefficients greater than 0.9 was observed between the bellows signals and displacement values calculated from the navigators. Sharpness of the portal vein of both conventional and silent navigator-triggered images was significantly higher than those of nontriggered images.
Conclusions:
The silent navigator-triggered zero-echo-time technique is feasible and might improve image quality and workflow of abdominal MRI of patients who are prone to acoustic noise. Magn Reson Med 79:2170-2175, 2018. © 2017 International Society for Magnetic Resonance in Medicine.
Related Concept Videos
Magnetic Resonance Imaging
Imaging Studies for Cardiovascular System IV: CMRI
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Radiological Investigation II: MRI and Ventilation Perfusion Scan
Magnetic Resonance Imaging (MRI) and Ventilation Perfusion Scans are two radiological investigations that offer detailed diagnostic images of the body, particularly lung structures.
MRI
MRI uses magnetic fields and radiofrequency signals to distinguish between normal and abnormal tissues. This technology provides a more detailed diagnostic image than CT scans, enabling it to characterize pulmonary nodules, stage bronchogenic carcinoma, and evaluate inflammatory activity in...
Imaging Studies IV: Magnetic Resonance Imaging