Related Experiment Video
Updated: Feb 25, 2026

Expired CO2 Measurement in Intubated or Spontaneously Breathing Patients from the Emergency Department
Published on: January 29, 2011
Optimization of Endotracheal Tube Cuff Pressure by Monitoring CO2 Levels in the Subglottic Space in Mechanically
Shai Efrati1, Gil Bolotin, Leon Levi
1From the *Assaf-Harofeh Medical Center, Zerifin, Israel; †Sackler School of Medicine and Sagol School of Neuroscience, Tel-Aviv University, Tel-Aviv, Israel; ‡Cardiac Surgery, Rambam Health Care Campus, Haifa, Israel; §Technion-Institute of Technology, Faculty of Medicine, Haifa, Israel; ∥Mayanei HaYeshua Medical Center, B'nai Brak, Israel; ¶Laniado Medical Center, Netanya, Israel; #Wolfson Medical Center, Holon, Israel; **Massachusetts General Hospital, Boston, Massachusetts; ††Harvard Medical School, Boston, Massachusetts; and ‡‡Ben-Gurion University of the Negev, Beer Sheva, Israel.
Background:
Many of the complications of mechanical ventilation are related to inappropriate endotracheal tube (ETT) cuff pressure. The aim of the current study was to evaluate the effectiveness of automatic cuff pressure closed-loop control in patients under prolonged intubation, where presence of carbon dioxide (CO2) in the subglottic space is used as an indicator for leaks. The primary outcome of the study is leakage around the cuff quantified using the area under the curve (AUC) of CO2 leakage over time.
Methods:
This was a multicenter, prospective, randomized controlled, noninferiority trial including intensive care unit patients. All patients were intubated with the AnapnoGuard ETT, which has an extra lumen used to monitor CO2 levels in the subglottic space.The study group was connected to the AnapnoGuard system operating with cuff control adjusted automatically based on subglottic CO2 (automatic group). The control group was connected to the AnapnoGuard system, while cuff pressure was managed manually using a manometer 3 times/d (manual group). The system recorded around cuff CO2 leakage in both groups.
Results:
Seventy-two patients were recruited and 64 included in the final analysis. The mean hourly around cuff CO2 leak (mm Hg AUC/h) was 0.22 ± 0.32 in the manual group and 0.09 ± 0.04 in the automatic group (P = .01) where the lower bound of the 1-sided 95% confidence interval was 0.05, demonstrating noninferiority (>-0.033). Additionally, the 2-sided 95% confidence interval was 0.010 to 0.196, showing superiority (>0.0) as well. Significant CO2 leakage (CO2 >2 mm Hg) was 0.027 ± 0.057 (mm Hg AUC/h) in the automatic group versus 0.296 ± 0.784 (mm Hg AUC/h) in the manual group (P = .025). In addition, cuff pressures were in the predefined safety range 97.6% of the time in the automatic group compared to 48.2% in the automatic group (P < .001).
Conclusions:
This study shows that the automatic cuff pressure group is not only noninferior but also superior compared to the manual cuff pressure group. Thus, the use of automatic cuff pressure control based on subglottic measurements of CO2 levels is an effective method for ETT cuff pressure optimization. The method is safe and can be easily utilized with any intubated patient.
More Related Videos
07:15Preoxygenation Techniques for Tracheal Intubation in Critically Ill Adults Utilizing Oxygen Mask and Noninvasive Ventilation
Published on: December 5, 2025
07:51Evaluation of Capnography Sampling Line Compatibility and Accuracy when Used with a Portable Capnography Monitor
Published on: September 29, 2020
Related Concept Videos
Assessment of Diffusion and Perfusion
The Role of Diffusion in Respiration
Diffusion is the process by which molecules move from an area of higher concentration to an area of lower concentration. In the respiratory system, this...
Endotracheal Intubation I: Procedure
The ET tube comprises various components, including a standard adaptor to attach a bag-valve-mask (BVM) or ventilator, a cuff, a pilot balloon, and radiopaque markings along its length to measure the insertion distance. The tube sizes...
Endotracheal Tube Extubation
Procedure
Extubation removes the endotracheal tube (ETT) from the patient on mechanical ventilation. It requires a well-coordinated, multidisciplinary approach involving physicians, nurses, respiratory therapists, and other healthcare professionals....
Endotracheal Intubation II: Nursing Management
1. Nursing Care of Patients Before Intubation
Before the endotracheal intubation procedure, nurses play an essential role in ensuring the process goes smoothly. The nurses must be familiar with intubation...
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...
Cardiopulmonary Resuscitation II: ACLS Airway Management