Related Experiment Video
Updated: Feb 25, 2026

Assessment of Boron Doped Diamond Electrode Quality and Application to In Situ Modification of Local pH by Water Electrolysis
Published on: January 6, 2016
Development of Conductive Boron-Doped Diamond Electrode: A microscopic, Spectroscopic, and Voltammetric Study
Kevin E Bennet1,2, Kendall H Lee3, James N Kruchowski4
1Division of Engineering, Mayo Clinic, Rochester, MN 55905, USA. bennet.kevin@mayo.edu.
Abstract:
Building on diamond characteristics such as hardness, chemical inertness and low electron emission threshold voltage, the current microscopic, spectroscopic and voltammetric investigations are directed towards improving the properties of electrode coating materials for their future use in clinical studies of deep brain stimulation via fast-scan cyclic voltammetry (FSCV). In this study we combine the capabilities of confocal Raman mapping in providing detailed and accurate analysis of local distributions of material constituents in a series of boron-doped polycrystalline diamond films grown by chemical vapor deposition, with information from the more conventional techniques of scanning electron microscopy (SEM) and infrared absorption spectroscopy. Although SEM images show a uniform distribution of film crystallites, they have the limitation of being unable to differentiate the distribution of boron in the diamond. Values of 1018-1021 atoms/cm³ of boron content have been estimated from the absorption coefficient of the 1290 cm-1 infrared absorption band and from the 500 cm-1 Raman vibration. The observed accumulation of boron atoms and carbon sp² impurities at the grain boundaries suggests that very high doping levels do not necessarily contribute to improvement of the material's conductivity, corroborating with voltammetric data. FSCV results also indicate an enhanced stability of analyte detection.
Related Concept Videos
Voltammetric Techniques: Cyclic Voltammetry
Voltammetric Techniques: Pulse Voltammetry
Potentiometry: Types of Electrodes
The Standard Hydrogen Electrode (SHE) is a widely used reference electrode that maintains zero potential across all temperatures. However, its need for a continuous hydrogen gas supply renders it impractical for everyday use.
An alternative to SHE is the Saturated Calomel Electrode (SCE). This electrode features an...
Voltammetric Techniques: Linear-Scan (E vs Time)
Voltammetry: Factors Affecting Measurements
Controlled-Potential Coulometry: Electrolytic Methods
The chosen potential...