Is the PTW 60019 microDiamond a suitable candidate for small field reference dosimetry?
Vanessa De Coste1, Paolo Francescon2, Marco Marinelli3
1Istituto Nazionale di Metrologia delle Radiazioni Ionizzanti, ENEA-INMRI, C R Casaccia, I-00123 Roma, Italy.
Physics in Medicine and Biology
|August 10, 2017
Summary
The PTW microDiamond dosimeter shows reproducible output factors in small fields, making it suitable for accurate small field reference dosimetry down to 5 mm. Monte Carlo simulations validate its performance.
Area of Science:
- Medical Physics
- Radiation Dosimetry
- Radiotherapy Physics
Background:
- Accurate dosimetry in small radiation fields is crucial for modern radiotherapy techniques.
- The PTW microDiamond (MD) dosimeter is investigated for its potential in small field reference dosimetry.
- Previous studies have shown varying results regarding the performance of diamond detectors in small fields.
Purpose of the Study:
- To systematically study the output factors (OF) of the PTW microDiamond (MD) dosimeter in small fields.
- To evaluate the suitability of the MD for small field reference dosimetry.
- To compare the MD's performance with PTW silicon diodes (Si-D) and validate Monte Carlo (MC) models.
Main Methods:
- Ten MDs were calibrated under 60Co irradiation.
- OF measurements were performed in 6 MV photon beams using CyberKnife M6, Varian DHX, and Elekta Synergy linacs.
- Monte Carlo (MC) simulations were used to calculate beam-quality correction factors ([Formula: see text]) and small field output correction factors ([Formula: see text]) for MD and Si-D.
Main Results:
- High reproducibility of dosimetric properties was observed among the ten MDs.
- Experimental [Formula: see text] values agreed within 1% with MC-calculated ones.
- Output correction factors were obtained within +0.7% and -1.4% down to 5 mm field sizes, showing good agreement with MC and Si-D results.
Conclusions:
- The PTW microDiamond dosimeter exhibits reproducible response and is validated by MC modeling.
- Accurate reference dosimetry is feasible using the MD for field sizes as small as 5 mm.
- The study confirms the MD's suitability for small field dosimetry applications in radiotherapy.