The (in-)validity of volatile POCT parameters from patients beyond normothermia
Michael A Gruber1, Stefan Felbermeir1, Regina Lindner1
1Department of Anesthesiology, Regensburg University Medical Center, Franz-Josef-Strauß-Allee 11, 93053 Regensburg, Germany.
Background:
A common characteristic of all blood gas analyzers on the market is that measurements are processed at 37°C, not at the real patients´ temperature. Subsequently temperature-sensitive parameters can be mathematically corrected (alpha-stat method) or used directly (pH-stat method). National rules in Germany (Rili-BAEK) demand defined accuracy and precision without any restriction to samples´ temperatures or corrections. As consequence in the investigation at hand we tried to find out whether blood gas analyzers can fulfill the regulations for pCO2 and pO2 when normothermia of the matrix is not given.
Methods:
Five matrices (blood from intensive care unit (ICU) patients, blood from healthy donors and 3 levels of bovine based quality control material) were tonometered at "high" and "low" partial pressures of O2 and CO2 within the RiLi-BAEK controlled range at 32, 37 and 40°C. One mL material was aspired into each blood gas (BG) syringe and analysis was accomplished immediately after. The procedure was repeated 10-fold for "high" and "low" gas concentrations and run on 4 different analyzers. At 18°C instead to the "high" one a "median" gas (n=10 as well) was employed. Every condition which constitutes of temperature (4), matrix (5), analyzer (4) and level of the partial pressure (2) led to a total of 1600 measurements.
Results:
At 32°C or 37°C matrix temperature 7.5% to 27.5% of the pCO2(T) and between 14.5% and 28.1% of the pO2(T) results were outside the borders required by the RiLi-BAEK. At 18°C or 40°C the number of results beyond the allowed borders grows up to 82.5% for pCO2(T) and 73% for pO2(T) depending on the partial pressure (PP) level.
Conclusions:
High precision in automated quality control (at a constant matrix temperature) is given in modern BGAnalyzers but is counteracted in practice by non normothermic patient's temperature and unavoidable sample handling effects.
More Related Videos
14:28Non-Invasive Monitoring of Microvascular Oxygenation and Reactive Hyperemia using Hybrid, Near-Infrared Diffuse Optical Spectroscopy for Critical Care
Published on: May 10, 2024
05:00Short-Duration Hypothermia Induction in Rats using Models for Studies examining Clinical Relevance and Mechanisms
Published on: March 3, 2021
Related Concept Videos
Factors Affecting Body Temperature
Factors may include:
Assessing Body Temperature - Rectal
Follow these steps for rectal temperature assessment:
Step 1: Perform hand hygiene and don clean gloves to prevent cross-infection.
Step 2: Position the patient in a side-lying position to better visualize the rectal...
Assessing Body Temperature - Temporal Artery
Step 1: Perform hand hygiene and don a fresh pair of gloves to prevent cross-infection and ensure patient safety.
Step 2: Explain the procedure to the patient to establish trust. Clear communication establishes trust with the patient, ensures they understand what to expect, promotes cooperation, and enhances comfort during the procedure.
Step 3: Assess the patient's...
Assessing Body Temperature - Oral
Step 1:
Start by practicing proper hand hygiene to prevent the spread of microorganisms.
Step 2:
Take the thermometer out of the charging unit, switch it on, and wait for the ready sign.
Step 3:
Gently slide the probe cover until a click is heard. This simple action prevents cross-contamination and ensures the correct placement of the probe cover.
Step 4:
Instruct the patient to open their mouth and place...
Homeostatic Imbalances in Body Temperature
Introduction to Vital Signs
Vital signs help healthcare professionals assess an individual's well-being and detect any functional changes...