Methacholine Challenge: Comparison of Airway Responsiveness Produced by a Vibrating Mesh Nebulizer Versus a Jet
Christianne M Blais1, Donald W Cockcroft1,2, Justine Veilleux3
11 Department of Physiology, University of Saskatchewan , Saskatoon, Saskatchewan, Canada .
Background:
The latest methacholine challenge testing (MCT) guidelines published by the European Respiratory Society recommend the characterization of nebulizers before their use in clinics and research. Such investigations are necessary for accurately determining the provocative dose of methacholine causing a 20% fall in FEV1 (PD20) delivered by a given device. The standard English Wright (Wright) jet nebulizer recommended in the 1999 guidelines by the American Thoracic Society has become difficult to obtain and possesses some characteristics that complicate the calculation of dose delivery from this device (e.g. evaporation). Our objective was to determine if the Aerogen® Solo (Solo) vibrating mesh nebulizer provides similar methacholine challenge test results compared to the currently used Wright jet nebulizer.
Methods:
Sixty mild-to-moderate asthmatics were studied across three research sites in a randomized crossover study. Both methacholine challenges were completed at least 24 hours apart within a 2-week period. Testing with the Wright device was performed as per the 2-minute tidal breathing protocol. The Solo study arm followed the same procedure except for a shorter inhalation time of 1 minute. The provocative concentration of methacholine causing a 20% fall in FEV1 (PC20) and the methacholine PD20 were calculated following each methacholine challenge.
Results:
The geometric mean methacholine PC20 values for the Solo and the Wright differed statistically (0.65 mg/mL vs. 2.58 mg/mL, respectively, p < 0.00001) and clinically. Between-nebulizer geometric mean methacholine PD20 results are comparable by clinical standards [81.7 μg (Solo) vs. 64.7 μg (Wright)], although the slight difference in dose was statistically significant (p = 0.018).
Conclusions:
The comparability of PD20 values between the Solo and the Wright validates the importance of reporting airway responsiveness to methacholine in terms of dose and not concentration, as stressed in the latest testing guidelines. This finding along with several benefits associated with the Solo make it a promising nebulizer for performing MCT.
Related Concept Videos
Inhaled Medications
Antiasthma Drugs: Methylxanthines
Theophylline is thought to inhibit phosphodiesterase enzymes, increasing intracellular levels of cyclic adenosine monophosphate (cAMP) and cyclic guanosine monophosphate (cGMP). This rise in cAMP and cGMP concentrations stimulates cardiac function,...
Factors Affecting Pulmonary Ventilation
Alveolar Surface Tension
The alveolar fluid lines the luminal surface of the alveoli and exerts a force called surface tension. This force is caused by the polar water molecules in the liquid being more strongly attracted to each...
Antiasthma Drugs: Muscarinic Receptor Antagonists
Antimuscarinic agents compete with ACh for the same binding site on the muscarinic receptors. By binding to these receptors, they inhibit the downstream effects of ACh and block the parasympathetic...
Pulmonary Function Tests
Pulmonary Function Tests are crucial diagnostic tools for assessing respiratory function, particularly in patients with chronic respiratory disorders. They comprehensively evaluate lung volumes, ventilatory function, breathing mechanics, diffusion, and gas exchange. These tests help diagnose pulmonary diseases and play a significant role in monitoring disease progression, evaluating disability, and assessing response to therapy.
PFTs involve using a spirometer, a...
Suctioning the Nasopharyngeal Airway
Equipment Required