Radiation dose with digital breast tomosynthesis compared to digital mammography: per-view analysis
Gisella Gennaro1, D Bernardi2, N Houssami3
1Radiology Unit, Veneto Institute of Oncology IOV- IRCCS, via Gattamelata, 64, 35128, Padua, Italy. gisella.gennaro@iov.veneto.it.
Digital breast tomosynthesis (DBT) delivers a modest increase in radiation dose compared to digital mammography (FFDM). This small increase should not hinder DBT adoption, especially when using synthetic 2D images.
Area of Science:
- Radiology
- Medical Imaging
- Radiation Oncology
Background:
- Digital breast tomosynthesis (DBT) is an advancement over digital mammography (FFDM).
- Concerns exist regarding the potential increase in radiation dose with DBT.
- Previous studies often compared DBT combined with FFDM versus FFDM alone.
Purpose of the Study:
- To directly compare the radiation dose delivered by FFDM and DBT for a single imaging view.
- To assess the mean glandular dose (MGD) for both FFDM and DBT examinations.
- To evaluate the impact of DBT on radiation exposure in breast cancer screening.
Main Methods:
- Utilized 4,780 FFDM and 4,798 DBT images from 1,208 women in a screening trial.
- Employed automatic software to determine volumetric breast density (VBD) from raw images.
- Calculated mean glandular dose (MGD) using Dance's model based on VBD and exposure data.
Main Results:
- Statistically significant differences in MGD were observed between FFDM and DBT for both CC and MLO views (p<0.0001).
- DBT resulted in a higher MGD compared to FFDM (CC: 1.858 mGy vs. 1.366 mGy; MLO: 1.877 mGy vs. 1.374 mGy).
- Bland-Altman analysis indicated an average 38% increase in DBT dose over FFDM, ranging from 0% to 75%.
Conclusions:
- Tomosynthesis shows a modest increase in breast radiation dose compared to FFDM.
- The radiation burden from DBT should not deter its use, especially with synthetic 2D imaging.
- Clinical benefits of DBT should be prioritized over minor radiation dose concerns.
More Related Videos
13:44Detection of Architectural Distortion in Prior Mammograms via Analysis of Oriented Patterns
Published on: August 30, 2013
07:57Positron Emission Tomography-based Dose Painting Radiation Therapy in a Glioblastoma Rat Model using the Small Animal Radiation Research Platform
Published on: March 24, 2022
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
X-ray Imaging