Intra-individual comparison between abdominal virtual mono-energetic spectral and conventional images using a novel
Jonas Doerner1, Christian Wybranski1, Jonathan Byrtus1
1Institute of Diagnostic and Interventional Radiology, University Hospital of Cologne, Cologne, Germany.
Objectives:
To quantitatively and qualitatively assess abdominal arterial and venous phase contrast-enhanced spectral detector computed tomography (SDCT) virtual mono-energetic (MonoE) datasets in comparison to conventional CT reconstructions provided by the same system.
Materials And Methods:
Conventional and MonoE images at 40-120 kilo-electron volt (keV) levels with a 10 keV increment as well as 160 and 200 keV were reconstructed in abdominal SDCT datasets of 55 patients. Attenuation, image noise, and contrast- / signal-to-noise ratios (CNR, SNR) of vessels and solid organs were compared between MonoE and conventional reconstructions. Two readers assessed contrast conditions, detail visualization, overall image quality and subjective image noise with both, fixed and adjustable window settings.
Results:
Attenuation, CNR and SNR of vessels and solid organs showed a stepwise increase from high to low keV reconstructions in both contrast phases while image noise stayed stable at low keV MonoE reconstruction levels. Highest levels were found at 40 keV MonoE reconstruction (p<0.001), respectively. Solid abdominal organs showed a stepwise decrease from low to high energy levels in regard to attenuation, CNR and SNR with significantly higher values at 40 and 50 keV, compared to conventional images. The 70 keV MonoE was comparable to conventional poly-energetic reconstruction (p≥0.99). Subjective analysis displayed best image quality for the 70 keV MonoE reconstruction level in both phases at fixed standard window presets and at 40 keV if window settings could be adjusted.
Conclusion:
SDCT derived low keV MonoE showed markedly increased CNR and SNR values due to constantly low image noise values over the whole energy spectrum from 40 to 200 keV.
More Related Videos
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies III: Computed Tomography
Imaging Studies for Cardiovascular System V: CT
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Imaging Studies for Cardiovascular System VI: Calcium -Scoring CT
Imaging Studies II: Ultrasonography