Related Experiment Video
Updated: Feb 24, 2026

An Implantable System For Chronic In Vivo Electromyography
Published on: April 21, 2020
Occurrence and extraction of implantable cardioverter-defibrillator leads with conductor externalization
Aleksander Maciąg1, Paweł Syska, Maciej Sterliński
1Institute of Cardiology Warsaw. maciag_o@poczta.onet.pl.
Background:
The increasing number of patients with implantable cardioverter-defibrillators (ICD) contributes to the rising number of patients qualifying for a transvenous lead extraction (TLE) due to infection, vascular or lead failure related indications. The purpose of this study was to perform a retrospective analysis of the occurrence of conductor externalization in TLE patients and to assess the success rate in the extraction of these leads.
Methods:
TLE procedure was performed between 2012 and 2014 of 428 electrodes in 259 patients. Out of these, 143 (33.4%) leads in 138 (52.9%) patients were ICD leads. The indications for the TLE in ICD patients were: infection in 37 patients, lead failure in 84 patients, and others in 17 patients. Conductor externalization was observed in 8 ICD leads (5.6%) in 8 (5.8%) patients. The mean dwell-ing time for externalized leads was 87.9 (55 to 132) months compared to 60.1 (3 to 246) months of the remaining 135 ICD leads (p = 0.0329). All externalized leads were successfully and completely extracted using device traction, mechanical telescopic sheaths and/or autorotational cutting sheaths. No complica-tions of lead extraction procedures were observed in 8 patients with externalization.
Results:
Patients with lead externalization were often in a better New York Heart Association func-tional class (I or II) compared to those in the rest of the study group (p = 0.0212).
Conclusions:
Conductor externalization is a rare finding in patients undergoing TLE. This occurs with different manufacturers and lead types. In this complication transvenous lead extraction with the mechanical extraction tools can be safely performed.
More Related Videos
12:45Benefits of Cardiac Resynchronization Therapy in an Asynchronous Heart Failure Model Induced by Left Bundle Branch Ablation and Rapid Pacing
Published on: December 11, 2017
16:40A New Single Chamber Implantable Defibrillator with Atrial Sensing: A Practical Demonstration of Sensing and Ease of Implantation
Published on: February 28, 2012