Oxygen beams for therapy: advanced biological treatment planning and experimental verification
O Sokol1, E Scifoni1,2, W Tinganelli1,2
1Biophysics Department, GSI Helmholtzzentrum für Schwerionenforschung GmbH, Planckstr. 1, D-64291 Darmstadt, Germany.
Abstract:
Nowadays there is a rising interest towards exploiting new therapeutical beams beyond carbon ions and protons. In particular, [Formula: see text]O ions are being widely discussed due to their increased LET distribution. In this contribution, we report on the first experimental verification of biologically optimized treatment plans, accounting for different biological effects, generated with the TRiP98 planning system with [Formula: see text]O beams, performed at HIT and GSI. This implies the measurements of 3D profiles of absorbed dose as well as several biological measurements. The latter includes the measurements of relative biological effectiveness along the range of linear energy transfer values from ≈20 up to ≈750 keV μ [Formula: see text], oxygen enhancement ratio values and the verification of the kill-painting approach, to overcome hypoxia, with a phantom imitating an unevenly oxygenated target. With the present implementation, our treatment planning system is able to perform a comparative analysis of different ions, according to any given condition of the target. For the particular cases of low target oxygenation, [Formula: see text]O ions demonstrate a higher peak-to-entrance dose ratio for the same cell killing in the target region compared to [Formula: see text]C ions. Based on this phenomenon, we performed a short computational analysis to reveal the potential range of treatment plans, where [Formula: see text]O can benefit over lighter modalities. It emerges that for more hypoxic target regions (partial oxygen pressure of ≈0.15% or lower) and relatively low doses (≈4 Gy or lower) the choice of [Formula: see text]O over [Formula: see text]C or [Formula: see text]He may be justified.
Related Concept Videos
Cancer Therapies
However, cancer treatments can pose several challenges, as therapies used to kill cancer cells are generally also toxic to normal cells. Moreover, cancer cells mutate rapidly and can develop resistance to chemical agents or radiation therapy. Besides, all types of cancer cells may not respond to the same therapy. Some cancer cells respond to one...
Treatment for Pulmonary Arterial Hypertension: Oxygen Therapy for Respiratory Failure
Oxygen therapy is vital in increasing and maintaining blood oxygen levels in PAH patients. As a result, it aids in reducing fatigue,...
Radiation: Applications
The average...
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...