Related Experiment Video
Updated: Feb 24, 2026

Assessing the Accuracy of Fitness Smartwatch Data for Cardiovascular and Physical Activity Monitoring: A Validation Study in Digital Health
Published on: February 21, 2025
Clinical Validation of Heart Rate Apps: Mixed-Methods Evaluation Study
Thijs Vandenberk1,2, Jelle Stans1, Christophe Mortelmans3
1Mobile Health Unit, Faculty of Medicine and Life Sciences, Hasselt University, Hasselt, Belgium.
Smartphone heart rate (HR) apps can be validated using beat-to-beat analysis against ECG. This method ensures accurate mobile HR monitoring, comparing smartphone photoplethysmography (PPG) technology with clinical standards.
Area of Science:
- Biomedical Engineering
- Digital Health
- Cardiovascular Technology
Background:
- Photoplethysmography (PPG) enables smartphone-based heart rate (HR) monitoring, offering scalable mobile health solutions.
- Current validation methods for HR apps lack standardization, leading to inconsistent accuracy assessments compared to clinical standards.
Purpose of the Study:
- To investigate and define essential elements for validating and comparing smartphone HR applications against established technologies.
- To establish a reliable methodology for assessing the accuracy of mobile HR monitoring solutions.
Main Methods:
- Two prospective studies utilized the FibriCheck app, comparing its HR measurements against FDA-cleared devices (Nonin oximeter, AliveCor ECG).
- A beat-to-beat interval analysis was performed, comparing FibriCheck's peak-to-peak intervals (PPI) with synchronized ECG R-R intervals (RRI).
Main Results:
- The first study showed high HR correlation (r=0.834-0.897) between FibriCheck and clinical devices, with no significant differences.
- The second study demonstrated a strong correlation (rs=0.993) and low error (RMSE=23.04 ms) between FibriCheck PPI and ECG RRI, with no significant difference.
Conclusions:
- Simultaneous HR measurement by a smartphone app and an ECG system, analyzed via beat-to-beat intervals, is the most suitable validation method.
- This approach enhances the accuracy of HR app assessments, promoting reliable mobile cardiovascular monitoring.
Related Concept Videos
Factors Influencing Heart Rate
Let us explore the significant factors affecting heart rate, including age, body temperature, posture, acute pain, chemical influences,...
Pulse rhythm
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac...
Dysrhythmias V: Evaluating Dysrhythmias
Assessment of blood pressure in brachial artery(two-step method)
Holter Monitor: 24-Hour Monitoring
Assessment of apical pulse
Assessing the apical pulse is a critical nursing procedure, particularly indicated for: