Related Experiment Video
Updated: Feb 24, 2026

Visualization of Low-Level Gamma Radiation Sources Using a Low-Cost, High-Sensitivity, Omnidirectional Compton Camera
Published on: January 30, 2020
Investigation of γ-γ coincidence counting using the National Nuclear Array (NANA) as a primary standard
S M Collins1, R Shearman1, J D Keightley2
1National Physical Laboratory, Hampton Road, Teddington, Middlesex TW11 0LW, United Kingdom; Department of Physics, University of Surrey, Guildford GU2 7XH, United Kingdom.
Abstract:
The National Physical Laboratory has recently been in the process of commissioning a multi-detector γ ray array - the National Nuclear Array (NANA). In this study we have sought to exploit the NANA and the excellent timing characteristics of its intrinsic LaBr3(Ce) scintillation detectors for use as a primary standardisation system. For this initial investigation, the absolute standardisation of 60Co has been performed by the γ-γ coincidence technique using NANA and the result compared to the established 4π(LS)-γ Digital Coincidence Counting (DCC) system. The effect of the angular correlation of the stretched E2 transitions emitted from the 4+→2+→0 states of 60Ni on the activity determined by NANA was observed between the pairs of detectors. Corrections for these angular correlations were derived through Monte Carlo simulations. An activity per unit mass by NANA of 330.8 (10) kBqg-1 for the 60Co solution was determined. There was no significant statistical difference between the results of NANA and the 4π(LS)-γ DCC, with a relative difference of 0.04% observed. This study shows that NANA can be used as a primary standard.
Related Concept Videos
Atomic Nuclei: Nuclear Magnetic Moment
Atomic Emission Spectroscopy: Instrumentation
Atomic Nuclei: Nuclear Spin State Population Distribution
Atomic Absorption Spectroscopy: Radiation and Light Sources
Two common narrow-range 'line' sources used in AAS are hollow-cathode lamps (HCLs) and...
Atomic Emission Spectroscopy: Overview
Atomic Absorption Spectroscopy: Instrumentation
The atomizer used in AAS can be either a flame atomizer or an...