Valved Holding Chambers and In Vitro Metered Dose Inhaler Performance: Effects of Flow Rate and Inhalation Delay
Xiaofei Liu1, Changning Guo1, Badrul Chowdhury2
11 Division of Pharmaceutical Analysis, Center for Drug Evaluation and Research , U.S. Food and Drug Administration, St. Louis, Missouri.
Background:
Multiple factors may influence the performance of a metered dose inhaler (MDI) when used with a valved holding chamber (VHC or "spacer").
Methods:
Andersen Cascade Impactor measurements were conducted for three MDI products and two different VHCs using a specially designed system that accommodated variable delay times between MDI actuation and introduction of the aerosol into the impactor, and allowed reduced flow through the VHC while the impactor was operated at 28.3 L/min. Deposited drug mass and aerodynamic particle size distribution were determined using validated high-performance liquid chromatography (HPLC) methods. A two-level, three-factor full-factorial design of experiments (DOE) design was applied to assess the influences of VHC type, flow rate, and inhalation delay on a total of seven performance characteristics for each MDI product. An experiment without a VHC was added to assess the influence of VHC presence.
Results:
DOE study shows the presence and type of VHC are the major influences on emitted dose and respirable fraction. Following the VHC effect, the inhalation delay has the most significant influence on most MDI performance metrics-emitted dose, respirable particle dose and fraction (aerosols between 1.1 and 4.7 μm), and fine particle dose and fraction (aerosols under 4.7 μm).
Conclusion:
This study illustrates the use of DOE analysis to effectively assess the effects of patient handling parameters (flow rate and inhalation delay) on the performance of MDI drugs when used with a VHC. The results of this study will inform Food and Drug Administration reviewers, the pharmaceutical industry, and healthcare practitioners as to safe and effective use of MDI products when used in conjunction with spacer devices.
More Related Videos
08:25Combining Volumetric Capnography And Barometric Plethysmography To Measure The Lung Structure-function Relationship
Published on: January 8, 2019
07:28Dry Powder and Nebulized Aerosol Inhalation of Pharmaceuticals Delivered to Mice Using a Nose-only Exposure System
Published on: April 6, 2017
Related Concept Videos
Inhaled Medications
Pipe Flowrate Measurement
The orifice meter is a simple,...
Factors Affecting Pulmonary Ventilation
Alveolar Surface Tension
The alveolar fluid lines the luminal surface of the alveoli and exerts a force called surface tension. This force is caused by the polar water molecules in the liquid being more strongly attracted to each...
Respiratory Volumes
Tidal Volume (TV) Tidal volume (TV) is the air inhaled or exhaled in a...
Mechanical Ventilation II: Invasive Ventilation
Negative-Pressure Ventilators
Negative-pressure ventilators create a vacuum around the chest or body to draw air into the lungs, simulating breathing. This method does not require an...
Assessment of Ventilation I: Respiratory Rate
A Ventilation assessment is critical for monitoring a patient's health status. Respiration, one of the most accessible vital signs, provides insights into the function of numerous body systems and can indicate serious health issues, such as brainstem injuries from head trauma.
Critical Guidelines for Assessing Ventilation: