Performance Testing Of Selected Types of Electronic Personal Dosimeters in X- and Gamma Radiation Fields
Nikola Kržanović1, Miloš Živanović, Olivera Ciraj-Bjelac
1*School of Electrical Engineering, University of Belgrade, 73 Bulevar kralja Aleksandra, 11120 Belgrade, Serbia; †Vinča Institute of Nuclear Sciences, Department of Radiation and Environmental Protection, Faculty of Physics, University of Belgrade, 12-14 Mike Petrovića Alasa, 11001 Vinča, Belgrade, Serbia; ‡Faculty of Physics, University of Belgrade, 12 Studentski trg, 11001 Belgrade, Serbia.
Abstract:
Electronic personal dosimeters (EPDs) are increasingly being used alongside conventional thermoluminescent dosimeters to measure the dose of legal record in terms of personal dose equivalent. Therefore, it is of great importance to execute performance tests of these dosimeters in photon fields of various energies and at various angles of incidence. This testing is done in order to simulate the behavior of these dosimeters in realistic multidirectional polyenergetic ionizing radiation fields. Tests of accuracy, linearity, energy response, and angular response have been performed on 10 EPDs from multiple manufacturers. Various radiation qualities have been used in the energy range from 33 keV to 1.33 MeV and for angles of incidence 0° to 80°. This research proves that many of the EPDs tested performed according to the manufacturer's specifications and the requirements of the international standards regarding personal dosimetry.
Related Concept Videos
Biological Effects of Radiation
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET
Gas Chromatography: Types of Detectors-II