Related Experiment Video
Updated: Feb 23, 2026

Improved UPLC-UV Method for the Quantification of Vitamin C in Lettuce Varieties Lactuca sativa L. and Crop Wild Relatives Lactuca spp.
Published on: June 30, 2020
Development and implementation of an HPLC-ECD method for analysis of vitamin C in plasma using single column and
Zlatuse D Clark1, Elizabeth L Frank2
1ARUP Institute for Clinical and Experimental Pathology, Salt Lake City, UT 84108, United States.
Objectives:
Vitamin C (l-ascorbic acid) is a water-soluble micronutrient necessary for human life. Inadequate intake can lead to the fatal disease scurvy. Measurement of vitamin C is used to assess nutritional status and to monitor supplementation. The goal of this study was to develop a chromatographic method for the quantitation of vitamin C in human plasma.
Design And Methods:
Samples were prepared by protein precipitation, addition of internal standard, and reduction with dithiothreitol. Separation of ascorbic acid was accomplished by isocratic elution on a reverse-phase column; concentration was determined by coulometry. The method was validated through studies of assay linearity, sensitivity, imprecision, accuracy, analytical specificity, and carryover.
Results:
The new assay was developed using a single pump/single analytical column HPLC system. Results correlated well with our previously used spectrophotometric method. The analytical measurement range was 1.0-2500 µmol/L. The injection-to-injection time was 13 min. Subsequently, to increase method throughput and shorten turnaround time, a dual LC pump system with a 2-position/10-port switching valve capable of performing automatic alternating column regeneration was validated and implemented. The injection-to-injection time was reduced 2-fold to 6 min. The method was linear to 5000 µmol/L; limit of quantification was 1.9 µmol/L. Total imprecision was less than 5%.
Conclusions:
We have developed a robust method suitable for routine clinical measurement of vitamin C in plasma specimens. The method incorporates a simplified sample preparation and a stable, non-endogenous internal standard to specifically quantify vitamin C. Faster throughput was achieved by employing an automatic alternating column regeneration system.
Related Concept Videos
High-Performance Liquid Chromatography: Instrumentation
Ion-Exchange Chromatography
High-Performance Liquid Chromatography: Introduction
In HPLC, two phases play a critical role in the separation process:
High-Performance Liquid Chromatography: Elution Process
Capillary Electrophoresis: Applications
Capillary zone electrophoresis (CZE) separates ionic components based on their electrophoretic mobility. It has been used to separate proteins, amino acids,...
Capillary Electrophoresis: Instrumentation