A pleural vacuum relief device for pleural drain unit use in the hyperbaric environment
Marco Gelsomino1,2, Theo Tsouras1, Ian Millar1
1Hyperbaric Medicine Service, The Alfred Hospital, Melbourne, Victoria, Australia.
A novel pleural vacuum relief (PVR) device prevents excessive negative intrapleural pressure during hyperbaric therapy. This PVR device enhances safety and allows faster pressure changes in pleural drain units (PDUs).
Area of Science:
- Medical Devices
- Hyperbaric Medicine
- Pulmonary Physiology
Background:
- Standard pleural drain units (PDUs) can induce excessive negative intrapleural pressure (IPP) and fluid reflux under hyperbaric conditions, posing risks.
- This can lead to significant patient morbidity and limit hyperbaric treatment protocols.
Purpose of the Study:
- To develop and evaluate a pleural vacuum relief (PVR) device to manage risks associated with PDUs under hyperbaric conditions.
- To assess the PVR device's ability to allow more rapid hyperbaric pressure change rates.
Main Methods:
- A custom PVR device with a one-way pressure relief valve and micro filter was designed.
- The PVR device was connected to a standard Pleur-Evac A-6000 intercostal drain.
- Simulated IPP was measured with and without the PVR device at various suction levels and compression rates.
Main Results:
- Without the PVR device, excessive negative IPP (>10 kPa·min⁻¹) caused backflow through the PDU water seal.
- With the PVR device, negative IPP remained clinically acceptable, enabling compression rates of at least 30 kPa·min⁻¹.
- The PVR device allowed suction settings up to -20 cmH₂O during all hyperbaric treatment phases.
Conclusions:
- The developed PDU PVR device effectively minimizes risks of excessive negative IPP during hyperbaric therapy.
- The device reduces attendant workload and allows for faster hyperbaric pressure changes.
- The PVR device should be used with suction for optimal performance.
More Related Videos
07:41Insertion, Maintenance, and Removal of the Percutaneous Dual Lumen Cannula Right Ventricular Assist Device
Published on: July 20, 2022
09:36Halogenated Agent Delivery in Porcine Model of Acute Respiratory Distress Syndrome via an Intensive Care Unit Type Device
Published on: September 24, 2020
Related Concept Videos
Ventilatory Modes
There are three ventilatory modes: full support, partial support, and spontaneous. These are described below.
Full Support Modes
Full support modes include controlled mechanical ventilation, continuous mandatory...
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...
Pneumothorax-II
Clinical Manifestations:
Mechanical Ventilation II: Invasive Ventilation
Negative-Pressure Ventilators
Negative-pressure ventilators create a vacuum around the chest or body to draw air into the lungs, simulating breathing. This method does not require an...
Oxygen Delivering System III: Tracheostomy and T-piece
Tracheostomy
A tracheostomy is a surgically created opening (stoma) in the anterior part of the trachea. It is used to establish a patient airway, bypass an upper airway obstruction, simplify the removal of secretions, permit long-term...
Endoscopic Studies II: Thoracocentesis
Description
Excess pleural fluid or air may accumulate in some respiratory disorders in the thoracic cavity. To treat pleural effusion, a physician conducts thoracentesis by carefully piercing the chest wall and entering...