Quantifying discrepancy between indicated and actual oxygen flow rates delivered by Comweld Ezi-flow low and standard
Yoav Aufgang1,2, Bridget Devaney1,3,4,5, Jason Chan1,2
1Department of Intensive Care and Hyperbaric Medicine, Alfred Health, Melbourne, Australia.
Introduction:
During clinical use of extracorporeal membrane oxygenation (ECMO) in hyperbaric conditions at our centre, upward titration of indicated sweep gas flow rates is required to maintain adequate CO₂ clearance. This project measured the impact of hyperbaric pressure on oxygen flow rates delivered by the Comweld Ezi-Flow flowmeters used in our centre.
Methods:
Oxygen flow rates through Comweld Ezi-flow standard and low oxygen gas flowmeters were set at 101.3 kPa (1 atmosphere absolute [atm abs]) and then measured at intervals up to 284 kPa (2.8 atm abs) using a calibrated gas flow analyser, with cross verification against a small Douglas bag test type apparatus. During testing, the chamber was compressed and decompressed at a rate of 10 kPa·min⁻¹. Flow rates during chamber compression and decompression were compared.
Results:
The indicated rate of oxygen gas flow through the unadjusted flowmeters changed minimally - typically rising by a maximum of half of the diameter of the indicator ball. The actual (volumetric) flow, tested across indicated flow rates from 3 to 12 L·min⁻¹, was consistently reduced by approximately 50% as the chamber pressure increased from 101.3 to 284 kPa (1 to 2.8 atm abs). A slightly smaller reduction was observed when assessing the low flowmeter across the same pressure range; reductions of 40.0 and 43.3% were demonstrated at 0.3 to 0.6 L·min⁻¹ respectively. Differences in flow rates between compression and decompression were minor except at the very lowest flows.
Conclusions:
At 284 kPa (2.8 atm abs), actual volumetric flow of oxygen through Comweld Ezi-Flow flowmeters is dramatically reduced and this needs appropriate compensation to ensure therapeutic aims are achieved.
More Related Videos
Related Concept Videos
Pipe Flowrate Measurement
The orifice meter is a simple,...
Pipe Flowrate Measurement: Problem Solving
Special considerations while measuring oxygen saturation
Ensuring accuracy in vital sign recordings while prioritizing patient comfort and minimizing anxiety is important.
Measurement of Fluid Pressure
A basic form of manometer is the piezometer, a vertical tube open at the top and filled with the same...
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...