Related Experiment Video
Updated: Feb 23, 2026

Cardiopulmonary Bypass in a Mouse Model: A Novel Approach
Published on: September 22, 2017
Clinical Comparative Study of Cardiopulmonary Bypass with Nikkiso and BioMedicus Centrifugal Pumps
Satoshi Ohtsubo1, Eiki Tayama1, David Short1
1Department of Surgery, Baylor College of Medicine, Houston, Texas, U.S.A.
Abstract:
The Nikkiso centrifugal pump was evaluated in elective adult open heard surgery in comparison with the BioMedicus pump. Ten patients using the Nikkiso pump (Group N), and 10 patients using the BioMedicus pump (Group B) were examined for (or to determine) hemato-biologic parameters and patient outcome data as well as pump controllability. During cardiopulmonary bypass (CPB), both pumps maintained systemic perfusion satisfactorily without any mechanical adverse event. Rotation speed of the Nikkiso centrifugal pump (3,580 ± 100 rpm) was significantly higher than that of the BioMedicus pump (3,170 ± 100 rpm; p < 0.05) whereas changes in free plasma hemoglobin, platelet count, blood urea nitrogen, and creatinine levels showed no significant differences between the two groups. Urine output in Group N for 30 min after the initiation of CPB (7.10 ± 1.50 ml/kg/h) was significantly higher than that in Group B (3.23 ± 0.46 ml/kg/h; p < 0.05). Patient outcome data were similar in both groups, such as duration of intensive care unit stay, hospital stay, postoperative intubation time, amount of postoperative bleeding, and amount of blood transfused. These equivalent results with the BioMedicus pump suggested that the Nikkiso pump can be used in open heart surgery as a reliable and atraumatic CPB pump.