Comparison of Centrifugal and Roller Pump Hemolysis Rates at Low Flow
Yuko S Moon1, Satoshi Ohtsubo1, Michael R Gomez1
1Departments of Pediatrics, Baylor College of Medicine, Houston, Texas, U.S.ASurgery, Baylor College of Medicine, Houston, Texas, U.S.A.
Artificial Organs
|September 5, 2017
Summary
A new centrifugal pump shows comparable hemolysis rates to traditional roller pumps during neonatal extracorporeal membrane oxygenation (ECMO) simulation. This centrifugal pump offers advantages in size and priming volume for simpler ECMO systems.
Area of Science:
- Biomedical Engineering
- Cardiovascular Devices
- Neonatal Critical Care
Background:
- Neonatal extracorporeal membrane oxygenation (ECMO) requires specialized pumps to manage blood flow.
- Conventional roller pumps are widely used but may cause significant hemolysis.
- A novel centrifugal pump has been developed for potential ECMO applications.
Purpose of the Study:
- To compare the in vitro hemolysis rates of a new centrifugal pump against a conventional roller pump.
- To evaluate pump performance under simulated neonatal ECMO conditions (low flow, high pressure).
- To assess the potential benefits of the centrifugal pump for neonatal ECMO systems.
Main Methods:
- In vitro comparison of hemolysis rates between a centrifugal pump and a Stöckert roller pump.
- Simulated neonatal ECMO conditions using flow rates of 0.3 L/min and 1 L/min at 200 mm Hg pressure.
- Statistical analysis to determine significant differences in hemolysis rates.
Main Results:
- No significant difference in hemolysis rates was observed between the centrifugal and roller pumps (p = 0.57).
- Hemolysis rates did not differ significantly between high and low flow conditions (p = 0.86).
- The centrifugal pump demonstrated comparable blood trauma to the roller pump under simulated neonatal ECMO conditions.
Conclusions:
- The tested centrifugal pump is suitable for neonatal ECMO, exhibiting no increased hemolysis compared to roller pumps.
- The centrifugal pump's smaller size and reduced priming volume (25 ml) offer advantages for developing more compact ECMO systems.
- This centrifugal pump may facilitate the creation of simpler and potentially more effective neonatal ECMO setups.