Less Blood Damage in the Impeller Centrifugal Pump: A Comparative Study with the Roller Pump in Open Heart Surgery
Tomohiro Nishinaka1, Hiroshi Nishida1, Masahiro Endo1
1Department of Cardiovascular Surgery, The Heart Institute of Japan, Tokyo Women's Medical College, Tokyo, Japan.
Insights
The Nikkiso Centrifugal Pump (NP) demonstrated reduced blood trauma during cardiopulmonary bypass (CPB) compared to roller pumps (RP). This centrifugal pump showed lower plasma-free hemoglobin and p-thromboglobulin levels, indicating improved biocompatibility in open-heart surgery.
Area of Science:
- Cardiovascular Surgery
- Biomedical Engineering
- Hematology
Background:
- Cardiopulmonary bypass (CPB) is essential for open-heart surgery.
- Traditional roller pumps (RP) can cause significant blood trauma.
- Novel centrifugal pumps aim to minimize CPB-induced complications.
Purpose of the Study:
- To evaluate the clinical performance of the Nikkiso Centrifugal Pump (NP), Model HPM15.
- To compare blood trauma markers between NP and RP during CPB.
- To assess hemodynamic performance and biocompatibility of the NP in open-heart surgery.
Main Methods:
- Prospective study involving 14 patients undergoing elective coronary artery bypass grafting.
- CPB was performed using the Nikkiso Centrifugal Pump (NP).
- Serum hemoglobin, platelet count, and p-thromboglobulin (pTG) levels were measured and compared to a historical RP group (n=10).
Main Results:
- No significant difference in CPB duration between NP (109 min) and RP (121 min) groups.
- Significantly lower serum p-thromboglobulin levels in the NP group at 90 min of CPB.
- Significantly lower plasma-free hemoglobin levels in the NP group at multiple time points post-CPB (p < 0.01).
- No significant difference in platelet depletion between the groups.
Conclusions:
- The Nikkiso Centrifugal Pump (HPM15) exhibits excellent hemodynamic performance.
- The NP demonstrates reduced blood trauma compared to roller pumps during CPB.
- The NP is a promising alternative for minimizing complications in open-heart surgery.
Abstract:
A centrifugal pump with an impeller (Nikkiso Centrifugal Pump, Model HPM15; Nikkiso Co. Ltd.) was applied to cardiopulmonary bypass (CPB) in 14 patients who underwent elective coronary artery bypass grafting. Serum hemoglobin level, platelet count, and serum p-thromboglobulin (pTG) level were measured during CPB. The results were compared with those obtained in a comparative roller pump (RP) group (n = 10). There was no difference in the time on CPB between the NP (109 min) and RP (121 min) groups. The serum pTG level (ng/ ml) was lower in the NP group than in the RP group (obtained 90 min after the initiation of CPB). The plasma-free hemoglobin level also was lower in the NP group than in the RP group (obtained 90 min after the initiation of CPB, 120 min after the initiation of CPB, immediately after the termination of CPB, 3 h after termination of CPB; p < 0.01). There was no significant difference in platelet depletion. The HPM15 pump showed excellent hemodynamic performance with less blood trauma compared with the roller pump in its clinical application to open heart surgery.