Less Blood Damage in the Impeller Centrifugal Pump: A Comparative Study with the Roller Pump in Open Heart Surgery

Tomohiro Nishinaka1, Hiroshi Nishida1, Masahiro Endo1

  • 1Department of Cardiovascular Surgery, The Heart Institute of Japan, Tokyo Women's Medical College, Tokyo, Japan.

Artificial Organs
|September 5, 2017
PubMed

Insights

The Nikkiso Centrifugal Pump (NP) demonstrated reduced blood trauma during cardiopulmonary bypass (CPB) compared to roller pumps (RP). This centrifugal pump showed lower plasma-free hemoglobin and p-thromboglobulin levels, indicating improved biocompatibility in open-heart surgery.

Area of Science:

  • Cardiovascular Surgery
  • Biomedical Engineering
  • Hematology

Background:

  • Cardiopulmonary bypass (CPB) is essential for open-heart surgery.
  • Traditional roller pumps (RP) can cause significant blood trauma.
  • Novel centrifugal pumps aim to minimize CPB-induced complications.

Purpose of the Study:

  • To evaluate the clinical performance of the Nikkiso Centrifugal Pump (NP), Model HPM15.
  • To compare blood trauma markers between NP and RP during CPB.
  • To assess hemodynamic performance and biocompatibility of the NP in open-heart surgery.

Main Methods:

  • Prospective study involving 14 patients undergoing elective coronary artery bypass grafting.
  • CPB was performed using the Nikkiso Centrifugal Pump (NP).
  • Serum hemoglobin, platelet count, and p-thromboglobulin (pTG) levels were measured and compared to a historical RP group (n=10).

Main Results:

  • No significant difference in CPB duration between NP (109 min) and RP (121 min) groups.
  • Significantly lower serum p-thromboglobulin levels in the NP group at 90 min of CPB.
  • Significantly lower plasma-free hemoglobin levels in the NP group at multiple time points post-CPB (p < 0.01).
  • No significant difference in platelet depletion between the groups.

Conclusions:

  • The Nikkiso Centrifugal Pump (HPM15) exhibits excellent hemodynamic performance.
  • The NP demonstrates reduced blood trauma compared to roller pumps during CPB.
  • The NP is a promising alternative for minimizing complications in open-heart surgery.