Film dosimetry using a smart device camera: a feasibility study for point dose measurements
Trent Aland1,2, Ekta Jhala3, Tanya Kairn2,4,5
1ICON integrated cancer centre, 9 McLennan Court, North Lakes QLD 4509, Australia.
Smartphone cameras can serve as a pseudo densitometer for film dosimetry. This low-cost system shows potential for accurate point dose measurements, especially where traditional equipment is inaccessible.
Area of Science:
- Medical Physics
- Radiation Dosimetry
- Imaging Technology
Background:
- Film dosimetry is crucial for radiation dose verification.
- Conventional dosimetry systems can be expensive and inaccessible.
- Smartphone imaging technology offers a potential low-cost alternative.
Purpose of the Study:
- To investigate the feasibility of using a smartphone camera as a film dosimetry system.
- To develop and evaluate a smartphone-based pseudo-densitometer.
- To assess the accuracy and reproducibility of smartphone dosimetry compared to a flatbed scanner.
Main Methods:
- A light-tight box and smartphone camera were used in reflective transmission mode.
- Gafchromic EBT3 films were irradiated to doses up to 16.89 Gy.
- Dose response curves were created using smartphone RGB channels and compared to a flatbed scanner.
Main Results:
- The smartphone system demonstrated a dose response curve for Gafchromic EBT3 film.
- The red colour channel yielded the best results with an uncertainty of 2.4% at 2.5 Gy.
- Smartphone dosimetry uncertainties were higher than the flatbed scanner but showed promise.
Conclusions:
- Smartphone-based film dosimetry is a viable proof of concept.
- This method offers a potentially affordable solution for point dose measurements.
- It is particularly useful in resource-limited settings or when conventional systems are unavailable.