Related Experiment Video
Updated: Feb 23, 2026

Detection of Architectural Distortion in Prior Mammograms via Analysis of Oriented Patterns
Published on: August 30, 2013
Large area alpha sources with a lip: Integral counting and spectral distortions
Lynne King1, Ryan Fitzgerald1, Ronald E Tosh1
1Physical Measurement Laboratory, National Institute of Standards and Technology, Gaithersburg, MD 20899-8642, USA.
None:
The detection efficiency for large area alpha sources with adjustable heights of a raised lip around the edge were measured by 2π gas-filled proportional counter. The variations in low-energy spectral shape were modeled using a Geant4 radiation and charge transport Monte Carlo simulation, to enable extrapolation of the spectrum to zero energy. COMSOL Multiphysics finite-element analysis was used to explore changes in the spectrum gain in the presence of a lip. It qualitatively reproduced an increase in peak height due to an increasing height of the source lip. A spectrum analysis procedure was developed to perform integral counting on sources with a lip. The experimental results were used to validate the model, which was then used to predict the changes in 2π counting efficiency for other source-lip geometries.
Related Concept Videos
Atomic Emission Spectroscopy: Interference
¹H NMR Signal Integration: Overview
Atomic Emission Spectroscopy: Overview
Atomic Emission Spectroscopy: Lab
Area Problem
Types of Radioactivity
Alpha (α) decay is the emission of an α particle from the nucleus. For example, polonium-210 undergoes α decay: